Micropilot M FMR 231 with HART/4...20 mA
Commissioning
E Hauser
57
6.5.1
Envelope curve with the ToF Tool
After the basic setup, an evaluation of the measurement using the envelope curve is
recommended.
L00-FMR2xxxx-20-00--00-en-008
!
Note!
If the level of echo is very weak or there is a heavy interference echo, an orientation of
the Micropilot can help optimise the measurement (increase of the useful echo/
reduction of the interference echo) (see "Orientation of the Micropilot" on Page 66).
When using the Wave Guide antenna
no
orientation is required!
6.5.2
User-specific applications (operation)
For details of setting the parameters of user-specific applications, see separate
documentation BA 221F/00/en "Description of the instrument functions for Micropilot M"