Micropilot FMR20 HART
Technical data
Hauser
65
Differing values in near-range applications
0
0.5 (1.64)
D[m (ft)]
∆
[mm (in)]
5 (0.2)
10 (0.39)
-30 (-1.18)
30 (1.18)
-5 (-0.2)
-10 (-0.39)
R
0.3 (0.98)
R
A0045305
51
Maximum measured error in near-range applications; values for standard version
∆
Maximum measured error
R
Reference point of the distance measurement
D
Distance from reference point of antenna
Measured value resolution
Dead band as per EN61298-2:
• Digital: 1 mm (0.04 in)
• Analog: 4 µA
Response time
The response time can be configured. The following step response times apply (in
accordance with DIN EN 61298-2) when damping is switched off:
Tank height
<10 m (32.8 ft)
Sampling rate
1 s
-1
Response time
<3 s
In accordance with DIN EN 61298-2, the step response time is the time following an
abrupt change in the input signal up until the changed output signal has adopted 90%
of the steady-state value for the first time.
Influence of ambient
temperature
The measurements are carried out in accordance with EN 61298-3
• Digital (HART,
Bluetooth
® wireless technology):
Standard version: average T
C
= ±3 mm (±0.12 in)/10 K
• Analog (current output):
• Zero point (4 mA): average T
K
= 0.02 %/10 K
• Span (20 mA): average T
K
= 0.05 %/10 K
Содержание Micropilot FMR20
Страница 79: ...Micropilot FMR20 HART Operating menu Endress Hauser 79 User interface 0 to 100 m ...
Страница 101: ......
Страница 102: ...www addresses endress com 71534964 71534964 ...