Maintenance
Hauser
33
2.
A0043015
Clean the sensor optics carefully using the cloth provided.
3.
A0043011
Screw on the new spot cap.
4. Calibrate sensor.→ 22
10.6 Checking the measurement function
1. Remove the sensor from the medium.
2. Clean and dry the spot cap.
3. After about 10 minutes, measure the oxygen saturation index in air (without
recalibration).
The measured value should be at 100 ± 2 % SAT.