Installation
Memosens COS81D
16
Hauser
Inclination angles other than those mentioned are not permitted. In order to avoid buildup
and condensation on the spot, do
not
install the COS81D-****U ***sensor upside down.
Follow the instructions for installing sensors in the Operating Instructions for the
assembly used.
5.1.2
Mounting location
1. Choose a mounting location that is easy to access.
2. Ensure that upright posts and assemblies are fully secured and vibration-free.
3. Choose a mounting location with an oxygen concentration that is typical for the
application.
5.2
Mounting the sensor
5.2.1
Measuring system
A complete measuring system comprises:
• a Memosens COS81D oxygen sensor
• Measuring cable CYK10
• A transmitter, e.g. Liquiline CM42, Liquiline CM44x/R, Liquiline CM44P, Liquiline
Compact CM72/82
• Optional: an assembly, e.g. permanent installation assembly CPA842, flow assembly or
retractable assembly CPA875
1
2
3
4
A0029064
5
Example of a measuring system with COS81D
1
Retractable assembly CPA875
2
Liquiline CM42 transmitter
3
Measuring cable CYK10
4
Memosens COS81D oxygen sensor