5.5 Programme/software
update using programme
diskette at the unit
1. Save the unit set up at the PC using the commands “Display/change unit
settings/create new unit/Set up data carrier (diskette/ATA flash card)” on a new
diskette.
2. Switch the unit off.
3. Place the new original programme diskette into the unit.
4. Simultaneously hold the second and third push button from the left when switching
the unit back on. Hold these until the green LED flashes.
5. Release the push buttons. The new programme will now be loaded (lasts for
approx. 5 minutes). The screen remains dark during the load phase, green LED
flashes. Acknowledge the RESET message with OK.
6. When required load the previously saved set up parameters back into the unit.
(“Diskette functions/ load set-up from diskette”).
Attention:
When updating a programme all settings found in the unit memory and on
diskette are deleted.
5. Set-up unit