8.6
Corrective maintenance on the complete measuring point
COM 2x3-DX/-DS
8.6.1
Oxygen transmitter DX/DS
Cleaning the front panel
To clean the front panel, only use standard
cleaning agents.The front panel is resistant to
the following cleaning agents (acc. to test
methode DIN 42 115):
• Isopropanol
• Thinned acids (3 %)
• Thinned alkalis (5 %)
• Ester
• Hydrocarbons
• Ketone
• Household cleaners
Caution!
Do not use any concentrated mineral
acids or alkalis, benzyl-alcohol, me-
thylene chloride or high-pressure wa-
ter vapour at over 100 °C.
Simulation of oxygen measurement:
The instrument versions COM 2x3-DX/-DS
work with the sensors COS 4 / COS 4HD, i.e.
with amperometric sensors without a preamp.
The temperature sensor is directly brought
out.
To perform a functional test of the instrument,
the COS 4 sensor can be simulated by resis-
tors. Since the display value greatly depends
on the sensor slope setting and on temperatu-
re, the following values should be considered
as reference values:
Simulation resistance
Display value
∞
(open)
0 mg O
2
/l
1.9 M
Ω
7 ... 13 mg O
2
/l
With 1.9 M
Ω
and a
temperature equivalent
resistance of 37.3 k
Ω
after calibration
9.0 ... 9.2 mg O
2
/l
4.06 M
Ω
3.3 ... 6.1 mg O
2
/l
Simulation of temperature sensor:
The COS 4 sensor uses an NTC sensor
(30.0 k
Ω
/25 °C) for temperature measurement.
Use the following equivalent resistances to
test the temperature measurement:
Simulation resistance
Display value
95.0 k
Ω
0.0 °C
58.7 k
Ω
10.0 °C
37.3 k
Ω
20.0 °C
30.0 k
Ω
25.0 °C
24.3 k
Ω
30.0 °C
Test procedure:
• Connect the O
2
equivalent resistance (e.g.
decade resistor) to terminals 90 and 91.
• Connect the temperature equivalent
resistance (e.g. decade resistor) to
terminals 11 and 12.
• Air pressure measurement:
On instruments equipped with the S
package, the air pressure measured can
be checked in field S11.
There is no air pressure measurement for
the DX/DS versions. Use the location
altitude setting to compensate for the air
pressure influence.
8.6.2
Oxygen sensor COS 4
Please refer to the operating instructions for
the COS 4 with regard to sensor maintenance
and troubleshooting!
The operating instructions contain detailed
information including:
• Sensor design and function
• Mounting and installation
• Electrical connection
• Start-up and calibration
• Calculation examples and tables for verifi-
cation of measured values
• Maintenance, regeneration, cleaning
• Troubleshooting table
• Accessories and spare parts
• Technical data and ordering information