Coding
Field
Selection or range
Factory setting
Display
Info
O
Function group
CURRENT
OUTPUT
Initial display in function group
CURRENT OUTPUT.
O1
Select current
output
Out1
Out2
Output 2 with appropriate version.
You can select a separate
characteristic for each output.
O2 (1)
Enter linear
characteristic
lin
=
linear
(1)
sim = Simulation (2)
Tab = Table (3)
The characteristic can have a
positive or negative slope.
O211
Selection of
current range
4-20
0–20
O212
0/4 mA value;
enter correspond-
ing O
2
or
temperature value
Version DX/DS:
0.00 mg/l
0.00 ... 20.00 mg/l
0.0 %SAT
0.0 ... 200.0 %SAT
0 hPa
0 ... 400 hPa
WX/WS with COS 3 /
3S/ 3HD / 31:
0.00 mg/l
0.00 ... 60.00 mg/l
0.0 % Sat
0.0 ... 600.0 % Sat
0 hPa
0 ... 1200 hPa
WX/WS with COS 71:
0.000 mg/l
0.000 ... 20.000 mg/l
0.0 % Sat
0.0 ... 200.0 % Sat
0 hPa
0 ... 400 hPa
0.0 C
-10.0 ... 60.0 C
Enter the O
2
or temperature
value here of the min. current
value (0/4 mA) at the transmitter
output.
Minimum amount between
0/4 mA and 20 mA value: see
field O213.
The factory settings are printed in
bold
; functions printed in
italics are not available in the basic version.
O
OUTPUT
O1
Sel.Out
Out1
O2
Sel.Type
lin
O211
Sel.Range
4-20
O212
0/4 mA
0.00
mg/l
Instrument configuration
Liquisys M COM 223 / 253
28
Hauser