Coding
Field
Selection or range
Factory setting
Display
Info
R211
Switch function of
R2 (1) off or on
Off
On
Settings made for the limit
contactor are not deleted by
switching the function off.
R212
Enter contact
switch-on point
Version DX/DS:
20.00 mg/l
0.00 ... 20.00 mg/l
200.0 %Sat
0.0 ... 200.0 %Sat
400 hPa
0...400 hPa
WX/WS with COS 3 / 3S
/ 3HD / 31:
20.00 mg/l
0.00...60 mg/l
200.0 % Sat
0.0...600.0 % Sat
400 hPa
0...1200 hPa
WX/WS with COS 71:
20.000 mg/l
0.000...20.000 mg/l
200.0 % Sat
0.0...200.0 % Sat
400 hPa
0...400 hPa
Never set the switch-on point and
the switch-off point to the same
value.
(The unit selected in A1 appears.)
R213
Enter contact
switch-off point
Version DX/DS:
20.00 mg/l
0.00 ... 20.00 mg/l
200.0 %Sat
0.0 ... 200.0 %Sat
400 hPa
0...400 hPa
WX/WS with COS 3 / 3S
/ 3HD / 31:
20.00 mg/l
0.00...60 mg/l
200.0 % Sat
0.0...600.0 % Sat
400 hPa
0...1200 hPa
WX/WS with COS 71:
20.000 mg/l
0.000...20.000 mg/l
200.0 % Sat
0.0...200.0 % Sat
400 hPa
0...400 hPa
Entering the switch-off point
selects either a max contact
(switch-off point
< switch-on point) or a min
contact (switch-off point
> switch-on point) and
implements a constantly required
hysteresis (see Fig. 5.4).
R214
Enter pick-up delay
0 s
0 ... 2000 s
R215
Enter dropout
delay
0 s
0 ... 2000 s
The factory settings are printed in
bold
; functions printed in
italics are not available in the basic version.
Off
R231
Function
20.00
mg/l
R212
On value
20.00
mg/l
R213
Off value
0
s
R214
On Delay
0
s
R215
Off Delay
OM253D5B.CHP
Liquisys M COM 223 / 253
Instrument configuration
Hauser
41