5.6
Service
Coding
Field
Selection or range
default (bold)
Display
Info
S
Function group
SERVICE
S1
Selection of
language
ENG
=
English
GER = German
FRA = French
ITA
= Italian
NEL = Dutch
ESP = Spanish
This field must be configured
once (see Chap. 5.1). Then you
can exit S1 and continue.
S2
Configure hold
S+C
=
During
configuration
and calibration
CAL = During
calibration
Setup = During
configuration
no
= No hold
S = Setup,
C = Calibration.
S3
Manual hold
Off
On
The setting remains active even
after a power failure.
S4
Enter hold dwell
period
10 s
0 ... 999 s
S5
Enter release
code for SW
upgrade to Plus
packet. A factory
setting for HS
version.
0
0 ... 9999
Release code see nameplate.
If the incorrect code is entered,
the system jumps back to the
Measurement menu. The number
is edited with the PLUS or MINUS
key and confirmed with the
ENTER key.
“1" is displayed in case of active
code.
S6
Enter release
code for SW
upgrade to
ChemoClean. A
factory setting for
-...15 version.
0
0 ... 9999
Release code see nameplate.
If the incorrect code is entered,
the system jumps back to the
Measurement menu. The number
is edited with the PLUS or MINUS
key and confirmed with the
ENTER key.
“1" is displayed in case of active
code.
S
SERVICE
ENG
S1
Language
S+C
S2
Auto HOLD
Aus
S3
Man. HOLD
10
S4
Cont.Time
s
0
S5
PlusCode
0
S6
CleanCode
The factory settings are printed in bold; functions printed in
italics are not available in the basic version.
Instrument configuration
Liquisys M COM 223 F / 253 F
44
Hauser