Coding
Field
Selection or range
default (bold)
Display
Info
R2311
Enter basic load
0%
0 ... 40%
Selecting the basic load, you
choose the desired dosage
quantity.
100% basic load corresponds to:
steadily on at R237 = on
F
max
at R237 = freq
20 mA at R237 = curr
R2 (4)
Configure
cleaning function
(timer)
LC PV =
limit contactor O
2
(1)
LC °C =
limit contactor T (2)
PID controller (3)
Timer
(4)
Clean = ChemoClean (5)
Cleaning is performed using only
one cleaning agent (usually
water); see Fig. 5.12.
Confirmation with ENTER
switches off a different, already
switched-on function and its
settings are reset to the default.
R241
Switch function of
R2 (4) off or on
Off
On
R242
Enter rinse /
cleaning time
30 s
0 ... 999 s
The hold and relay settings are
activated for the period of time
specified here.
R243
Enter pause time
360 min
1 ... 7200 min
The pause time is the time
between two cleaning cycles
(see Chap. 5.5.3).
R244
Enter minimum
pause time
120 min
1 ... 3600 min
The minimum pause time
prevents continuous cleaning
when the cleaning trigger is
present.
R2 (5)
Configure
cleaning with
ChemoClean
(for four-contact
version and
assignment of
contacts 3 and 4)
GW PW = limit
contactor O
2
(1)
GW °C = limit
contactor T (2)
PID controller (3)
Timer (4)
Clean
=
ChemoClean
(5)
see Chap. 5.5.4.
Contact 3 = Water,
Contact 4 = Cleaner.
Confirmation with ENTER
switches off a different, already
switched-on function and its
settings are reset to the default.
R251
Switch function of
R2 (5) off or on
Off
On
The factory settings are printed in bold; functions printed in
italics are not available in the basic version.
R2411
%
BasicLoad
0
Timer
R2
Sel. Type
Off
R241
Function
30
R242
RinseTime
s
360
R243
PauseTime
min
120
R244
Min.Pause
min
Clean
R2
Sel. Type
Off
R251
Function
Instrument configuration
Liquisys M COM 223 F / 253 F
42
Hauser