Description of Device Parameters
Liquipoint FTW33
38
Hauser
Description
To prevent rapid switching back and forth when values are close to the switch point
"SP1"/"SP2" or switchback point "rP1"/"rP2", a delay in the range of 0.3 to 60 seconds, to one
decimal place, can be set for individual points.
If the measured value leaves the switching range during the delay time, the delay time
starts again from scratch.
rP1
SP1
OU1
%
HNO (MIN)
A
1
0
dS1
dR1
dS1
rP2
SP2
OU2
HNC (MAX)
t
1
0
A0034590
0
0-signal, output open in the quiescent state
1
1-signal, output closed in the quiescent state
A
Hysteresis (difference between the value of the switch point "SP1" and the value of the switchback point "rP1")
HNO Normally open contact (MIN)
HNC Normally closed contact (MAX)
%
Coverage of sensor
SP1 Switch point 1/SP2: Switch point 2
rP1 Switchback point 1/rP2: Switchback point 2
dS1 Set time for which the specific switch point must be reached continuously without interruption until an
electrical signal change takes place.
dR1 Set time for which the specific switchback point must be reached continuously without interruption until an
electrical signal change takes place.
Switch on value
Last value selected prior to switching off.
Options
No selection. The user is free to edit the values.
Input range
0.3 to 60 s
Factory setting
0.5 s (Switching delay time dS1/dS2)
1.0 s (Switchback delay time dR1/dR2)
Output 1/2 (OU1/OU2)
Navigation
Application → Output Switch 1/2 → Output 1/2 (OU1/OU2)
Description
Hysteresis: Determining whether sensor is free or covered.
Switch on value
Last function selected prior to switching off.
Options
• Hysteresis normally open (MIN)
• Hysteresis normally closed (MAX)