Liquiline System CA80TN
Operation
Hauser
65
Menu/Setup/Analyzer/Measurement
Function
Options
Info
Digestion time
0:15 to 2:00 (
hh:mm
)
Factory setting
0:30
Digestion temperature
100 to 120 °C
Factory setting
120 °C
Signal delay
0 to 600 s
Factory setting
0 s
Measurement active
: You can delay the process
of outputting the signal during an active
measurement by the defined time. Measuring is
paused for the duration of the signal delay time.
10.3.3 Reference measurement
Menu only available for device versions CA80TN-HR.
Menu/Setup/Analyzer/Reference sample inspection
Function
Options
Info
Start condition
Selection
• Immediate
• Date/time
• Disabled
Factory setting
Immediate
Immediate
: The reference measurement starts
immediately once the system changes to the
automatic mode.
Date/time
: The reference measurement starts at
the set time.
Disabled
: No reference measurement
Date
01.01.1970 to 07.02.2106
Functions only available for:
Start condition
=
Date/time
Time
00:00:00 to 23:59:59
Inspection interval
0-12 to 1-00 (
DD-hh
)
Factory setting
1-00
DD-hh
Function only available for:
Start condition
=
Immediate
or
Date/time
Setting the test interval
Nominal concentration
1 to 200 mg/l N
Factory setting
10 mg/l N
Concentration of the reference sample
Acceptance limit
Read only
10.3.4 Calibration
Menu/Setup/Analyzer/Calibration
Function
Options
Info
Start condition
Selection
• Immediate
• Date/time
• Disabled
Factory setting
Immediate
The calibration can start either immediately or at
the set time.
If the start condition selected is
Immediate
Date
01.01.1970 to 07.02.2106
Functions only available for:
Start condition
=
Date/time
Time
00:00:00 to 23:59:59
Calibration interval
0-01 to 90-00 (
DD-hh
)
Factory setting
04-00
DD-hh
For configuring the time interval of the
calibration/adjustment.