Diagnostics and troubleshooting
Liquiline System CA80TN
128
Hauser
11.13 Device test
11.13.1 Analyzer
DIAG/System testAnalyzer/Photometer
Function
Options
Info
Photometer status
Read only
Displays the status of the photometer
• 100 to 70%
Good state
• 70 to 50%
• Cuvette somewhat dirty
• Lamp or filter starting to age
• Measurement is OK, continue monitoring
• 50 to 20%
• Cuvette dirty
• Lamp or filter aging
• Can affect the measurement
• < 20%
• Maintenance needed: reactor, filter or lamp
• Reliable measurement is no longer possible
Number of flashes
Read only
Failed flashes
Read only
Start photometer status re-evaluation
Intensity measurement with zero standard to
reevaluate the photometer status
Photometer test
The test determines the following values:
• Current intensity 1
• Current intensity 2
• Lamp Voltage
• Temperature
Start for 5 min
DIAG/System testAnalyzer/Cooling module
Function
Options
Info
Temperature (reagent)
Read only
Current consumption
Read only
Temperature (extern)
Read only
Mode
Read only
Cooling for 60 minutes
Off for 60 minutes
1)
Test the cooling function of the Peltier cooler
DIAG/System testAnalyzer/Reactor
Function
Options
Info
Reactor temperature
Read only
Displays the temperature of the digestion reactor
Reactor current
Read only
Displays the reactor current
Reactor state
Read only
Reactor fan
Read only
On