Liquiline System CA80TN
Maintenance
Hauser
133
Rinsing the system
Measurement, cleaning or calibration is not possible while the bottles are removed.
Emptying the hoses
1. Stop the sample supply.
2. If you have not already done so,
set the manual mode.
3. Pull all the hoses out of the reagent bottles as far as they will go so that the hose
ends are outside of the liquid.
4. In addition, disconnect the sample hose (P) and remove the hose of the dilution
water (W) from the bottle so that the hose end is outside of the liquid.
5. Select
Menu
/
Operation
/
Maintenance
/
Decommissioning
/
Empty hoses
to remove
the liquid from the hoses.
The hoses are automatically purged with air and emptied. This process takes
several minutes.
6. Wait until this action has finished. Under
MODE
/
Info
you can check whether the
action has finished.
Rinsing the analyzer with water
1. Open all the reagent bottles and carefully remove the hoses together with the covers.
Then close the bottles firmly.
2. Dry the ends of the hoses with a clean paper towel.
3. Immerse all the hoses into a beaker with approx. 200 ml of distilled water in order to
rinse the system with water.
4. Trigger the action:
Menu
/
Operation
/
Maintenance
/
Decommissioning
/
Rinse with
water
.
The system is rinsed automatically and emptied. This process takes approx. 8
minutes.
5. Once the action is finished, remove the hoses from the beaker and dry them with a
clean paper towel.
6. To remove the liquid from the hoses:
Menu
/
Operation
/
Maintenance
/
Decommissioning
/
Empty hoses
.
The hoses are automatically purged with air and emptied. This process takes
several minutes.
7. Wait until this action has finished. Check whether the action has finished:
MODE
/
Info
.
8. Put all the hoses back into the bottles and connect the sample hose to the valve.
Cleaning the reactor
Flushing with water
1.
Menu
/
Operation
/
Maintenance
/
Reactor maintenance
/
Flushing reactor
The reactor is flushed with water 5 times.
2. Perform photometer test:
DIAG
/
System test
/
Analyzer
/
Photometer
/
Photometer
test
/
Start for 5 min
.
3.
Current intensity 1
: Note down the value.
The value varies. Rounding to the thousand value is accurate enough. The
hundreds are not significant. Example: 18500, 18734, 18494, 18611, 18342 →
note down "18xxx".