Endress+Hauser Levelflex FMP52 Скачать руководство пользователя страница 81

Levelflex FMP55 HART

Diagnostics and troubleshooting

Hauser

81

12.2.2  Calling up remedial measures

X X X X

X X X

X

X

X X X X

X X X

X

X

S

S

XX

20.50

X

i

S801

Menu

S

(ID:203)

S801   0d00h02m25s

1

2

4

6

3

5

Increase supply voltage

S801 Supply voltage

Diagnostic list

Diagnostics 1

Diagnostics 2
Diagnostics 3

Supply voltage

Supply voltage

  A0013940-EN

 22

Message for remedial measures

1

Diagnostic information

2

Short text

3

Service ID

4

Diagnostic behavior with diagnostic code

5

Operation time of occurrence

6

Remedial measures

The user is in the diagnostic message.

1. Press  (  symbol).

Diagnostic list

 submenu opens.

2. Select the desired diagnostic event with  or  and press  .

 The message for the remedial measures for the selected diagnostic event opens.

3. Press  +  simultaneously.

 The message for the remedial measures closes.

The user is in the 

Diagnostics

 menu at an entry for a diagnostics event, e.g. in 

Diagnostic

list

 submenu or in 

Previous diagnostics

.

1. Press .

 The message for the remedial measures for the selected diagnostic event opens.

2. Press  +  simultaneously.

 The message for the remedial measures closes.

Содержание Levelflex FMP52

Страница 1: ...Products Solutions Services Operating Instructions Levelflex FMP55 HART Guided wave radar BA01003F 00 EN 18 16 71344071 01 03 zz Device firmware...

Страница 2: ...2 Endress Hauser Order code Ext ord cd Ser no www endress com deviceviewer Endress Hauser Operations App XXXXXXXXXXXX XXXXX XXXXXX XXX XXXX XX Serial number 1 3 2 A0023555...

Страница 3: ...ice 30 6 2 1 Required mounting tools 30 6 2 2 Mounting the device 30 6 2 3 Mounting the Sensor remote version 30 6 2 4 Turning the transmitter housing 32 6 2 5 Turning the display module 33 6 3 Post i...

Страница 4: ...story 85 12 6 2 Filtering the event logbook 85 12 6 3 Overview of information events 85 12 7 Firmware history 87 13 Maintenance 88 13 1 Exterior cleaning 88 13 2 Cleaning coax probes 88 14 Repairs 89...

Страница 5: ...avoid this situation can result in serious or fatal injury CAUTION CAUTION This symbol alerts you to a dangerous situation Failure to avoid this situation can result in minor or medium injury NOTICE...

Страница 6: ...res processes or actions that are permitted Preferred Procedures processes or actions that are preferred Forbidden Procedures processes or actions that are forbidden Tip Indicates additional informati...

Страница 7: ...rea non hazardous area Indicates the non hazardous area 1 2 6 Symbols at the device Symbol Meaning Safety instructions Observe the safety instructions contained in the associated Operating Instruction...

Страница 8: ...ndividual parameter in the operating menu The description is aimed at those who work with the device over the entire life cycle and perform specific configurations Special documentation SD00326F Funct...

Страница 9: ...P55 XA00532F XA00532F XA00532F XA00572F XA00532F GA EAC Ex ia IIC T6 Ga FMP55 XA01380F XA01380F XA01380F XA01381F XA01380F GB EAC Ex ia IIC T6 Ga Gb FMP55 XA01380F XA01380F XA01380F XA01381F XA01380F...

Страница 10: ...66 FMP55 XA00639F XA00639F XA00639F XA00645F XA00639F 8A FM CSA IS XP Cl I II III Div 1 Gr A G FMP55 XA00531F XA00532F XA00531F XA00532F XA00531F XA00532F XA00572F XA00573F XA00531F XA00532F 1 A 2 wir...

Страница 11: ...to the following table 1 Feature 010 Approval Feature 030 Display Operation Ex marking BG L or M ATEX II 3G Ex nA ia Ga IIC T6 Gc BH L or M ATEX II 3G Ex ic ia Ga IIC T6 Gc B3 L or M ATEX II 1 2G Ex d...

Страница 12: ...data and listed in the Operating Instructions and supplementary documentation the measuring device may be used for the following measurements only Measured process variable Level and or interface Cal...

Страница 13: ...d in the hazardous area e g explosion protection pressure vessel safety Based on the nameplate check whether the ordered device is permitted for the intended use in the hazardous area Observe the spec...

Страница 14: ...3 Product description 3 1 Product design 3 1 1 Levelflex FMP51 FMP52 FMP54 FMP55 1 2 5 3 4 6 A0012399 1 Design of the Levelflex 1 Electronics housing 2 Process connection here as an example flange 3...

Страница 15: ...7 8 A0012422 2 Design of the electronics housing 1 Electronics compartment cover 2 Display module 3 Main electronics module 4 Cable glands 1 or 2 depending on instrument version 5 Nameplate 6 I O elec...

Страница 16: ...ks HART Registered trademark of the FieldComm Group Austin USA KALREZ VITON Registered trademark of DuPont Performance Elastomers L L C Wilmington USA TEFLON Registered trademark of E I DuPont de Nemo...

Страница 17: ...nt If required see nameplate Are the Safety Instructions XA present If one of the conditions does not comply contact your Endress Hauser distributor 4 2 Product identification The following options ar...

Страница 18: ...ment number of the Safety Instructions e g XA ZD ZE 12 2 D matrix code QR code 13 Modification mark 14 Manufacturing date year month 15 Permitted temperature range for cable 16 Ger terevision Dev Rev...

Страница 19: ...or the probe but at the process connection Take into account the mass center of the device in order to avoid unintended tilting Comply with the safety instructions transport conditions for devices ove...

Страница 20: ...Storage Transport Levelflex FMP55 HART 20 Endress Hauser 1 2 4 3 A0015471...

Страница 21: ...P55 Rod probes rope probes must be mounted in a stilling well or bypass 26 Coax probes can be mounted at an arbitrary distance from the wall of the vessel When mounting in the open a weather protectio...

Страница 22: ...te cable as ordered Product structure feature 600 Probe Design Option MB Sensor remote 3m 9ft cable The remote cable is supplied with these device versions Minimum bending radius 100 mm 4 inch A mount...

Страница 23: ...ensor Feature 060 Probe Tensile load limit kN FMP55 NA ND Rope 4mm 1 6 PFA 316 2 Bending strength of rod probes Sensor Feature 060 Probe Bending strength Nm FMP55 CA CB Rod 16mm 0 63 PFA 316L 30 Bendi...

Страница 24: ...FE flange cladding also serves as a seal between the nozzle and the device flange Flange size Number of screws Recommended torque Nm minimum maximum EN DN40 PN40 4 35 55 DN50 PN16 4 45 65 DN50 PN40 4...

Страница 25: ...HART Mounting Endress Hauser 25 6 1 5 Securing the probe Securing coax probes For WHG approvals For probe lengths 3 m 10 ft a support is required A0012608 Coax probes can be supported at any point of...

Страница 26: ...e below Minimum distance between end of probe and lower edge of the bypass Type of probe Minimum distance Rope 10 mm 0 4 in Rod 10 mm 0 4 in Coax 10 mm 0 4 in Pipe diameter 40 mm 1 6 for rod probes Ro...

Страница 27: ...th condensate formation water and a medium with low dielectric constant e g hydrocarbons In the course of time the bypass is filled with condensate up to the lower disposal and for low levels the the...

Страница 28: ...28 Endress Hauser Underground tanks A0014142 Use a coax probe for nozzles with large diameters in order to avoid reflections at the nozzle wall Non metallic vessels When mounting Levelflex in a non me...

Страница 29: ...ded in normal tank insulation to prevent the electronics heating up as a result of heat radiation or convection The insulation may not exceed beyond the points labeled MAX in the drawings 40 1 57 40 1...

Страница 30: ...be into the vessel observe the following Uncoil rope and lower it slowly and carefully into the vessel Do not kink the rope Avoid any backlash since this might damage the probe or the vessel fittings...

Страница 31: ...Probe electronics and connection cable are adjusted to match each other They are marked by a common serial number Only components with the same serial number shall be connected to each other If the me...

Страница 32: ...plug at the electronics housing C Length of the remote cable as ordered 6 2 4 Turning the transmitter housing To provide easier access to the connection compartment or display module the transmitter h...

Страница 33: ...rom the transmitter housing 3 Pull out the display module with a gentle rotation movement 4 Rotate the display module into the desired position Max 8 45 in each direction 5 Feed the spiral cable into...

Страница 34: ...example Process temperature Process pressure refer to the chapter on Material load curves of the Technical Information document Ambient temperature range Measuring range m Are the measuring point iden...

Страница 35: ...Terminal assignment 2 wire 4 20mA HART A Without integrated overvoltage protection B With integrated overvoltage protection 1 Active barrier with power supply e g RN221N Observe terminal voltage 2 HAR...

Страница 36: ...ed overvoltage protection B With integrated overvoltage protection 1 Active barrier with power supply e g RN221N Observe terminal voltage 2 HART communication resistor 250 Observe maximum load 3 Conne...

Страница 37: ...nt output 1 3 Supply voltage for current output 1 e g RN221N Observe terminal voltage 4 Cable screen observe cable specification 5 HART communication resistor 250 Observe maximum load 6 Connection for...

Страница 38: ...bserve maximum load 3 Connection for Commubox FXA195 or FieldXpert SFX350 SFX370 via VIATOR Bluetooth modem 4 Analog display device Observe maximum load 5 Signal cable including screening if required...

Страница 39: ...0 Supply voltage Observe terminal voltage observe cable specification 11 Terminal for potential equalization 12 Cable entry for signal line 13 Cable entry for power supply LCAUTION To ensure electrica...

Страница 40: ...d state relay Phoenix Contact OV 24DC 480AC 5 with mounting rail connector UMK 1 OM R AMS Electromechanical relay Phoenix Contact PLC RSC 12DC 21 3 2 1 4 A0015910 14 Connection of a digital input 1 Pu...

Страница 41: ...puts connected as power source The HART loop converter HMX50 can be acquired using the order number 71063562 Additional documentation TI00429F and BA00371F 7 1 2 Cable specification Devices without in...

Страница 42: ...bus plug connector M12 or 7 8 the signal line can be connected without opening the housing Pin assignment of the M12 plug connector 2 1 3 4 A0011175 Pin Meaning 1 Signal 2 not connected 3 Signal 4 Gro...

Страница 43: ...A0014080 1 Feature 020 of the product structure option A 2 Feature 010 of the product structure 2 wire 4 20 mA HART switch output 1 Approval 2 Terminal voltage U at the device Maximum load R dependin...

Страница 44: ...0 at the supply unit any Channel 1 18 5 to 30 V R W V U0 10 18 5 29 5 20 30 0 500 A0014080 Channel 2 12 to 30 V R W V U0 10 12 23 20 30 0 500 A0022583 1 Feature 020 of the product structure option C 2...

Страница 45: ...610 Accessory mounted option NA Overvoltage protection Technical data Resistance per channel 2 0 5 max Threshold DC voltage 400 to 700 V Threshold impulse voltage 800 V Capacitance at 1 MHz 1 5 pF No...

Страница 46: ...t the cable in accordance with the terminal assignment 35 8 When using screened cable Connect the cable screen to the ground terminal 9 Screw the cover onto the connection compartment 10 For instrumen...

Страница 47: ...Are all cable glands installed firmly tightened and correctly sealed m Does the supply voltage match the specifications on the transmitter nameplate m Is the terminal assignment correct 35 m If requir...

Страница 48: ...erature for the display 20 to 70 C 4 to 158 F The readability of the display may be impaired at temperatures outside the temperature range Operating elements local operation with 3 push buttons extern...

Страница 49: ...through the glass of the cover 8 1 3 Remote operation Via HART protocol 1 4 5 7 9 6 8 2 3 A0013764 17 Options for remote operation via HART protocol 1 PLC programmable logic controller 2 Transmitter...

Страница 50: ...lflex FMP55 HART 50 Endress Hauser Via service interface CDI E 1 2 3 A0014019 1 Service interface CDI of the measuring device Endress Hauser Common Data Interface 2 Commubox FXA291 3 Computer with Fie...

Страница 51: ...ains the last 20 messages which are no longer active Device information Contains information needed to identify the device Measured values Contains all current measured values Data logging Contains th...

Страница 52: ...nd analyze operational errors 1 In case of operation via operating tools e g FieldCare the Language parameter is located at Setup Advanced setup Display 2 only for operation via a FDT DTM system 3 onl...

Страница 53: ...from unauthorized access 54 Access authorization to parameters User role Read access Write access Without access code from the factory With access code Without access code from the factory With acces...

Страница 54: ...ministration Define access code 2 Define a max 4 digit numeric code as an access code Write protection is active Parameters that can always be changed The write protection does not include certain par...

Страница 55: ...code appears 2 Enter the access code The symbol in front of the parameters disappears all previously write protected parameters are now re enabled 8 2 5 Deactivation of the write protection via access...

Страница 56: ...s are now read only and cannot be edited any more exception Contrast display parameter Via local display Via service interface CDI Via HART protocol E 3 mm XXXXXXXXXXXXX WP CDI OFF ON A0026157 1 Loose...

Страница 57: ...this the symbol appears in the header of the measured value display and in the navigation view in front of the parameters XXXX XXX X X 20 50 XX XX A0015870 If the hardware write protection is disabled...

Страница 58: ...ing the keylock For the SD03 display The keylock is automatically activated If the device has not been operated via the display for 1 minute After a restart of the device To activate the keylock manua...

Страница 59: ...l if an error is active 1 2 Measured value symbols 1 3 Measured value 1 4 Unit 2 Measured value display 1 bargraph 1 value 2 1 Bargraph for measured value 1 2 2 Measured value 1 including unit 2 3 Mea...

Страница 60: ...e Diagnostics menu Status signals A0013956 Failure A device error is present The measured value is no longer valid A0013959 Function check The device is in service mode e g during a simulation A001395...

Страница 61: ...106 Terminal voltage A0012104 Temperature of the electronics or the sensor Measuring channels A0012000 Measuring channel 1 A0012107 Measuring channel 2 Status of the measured value A0012102 Status Ala...

Страница 62: ...he parameter For text and numeric editor Pressing the key briefly Opens the selected group Carries out the selected action Pressing the key for 2 s confirms the edited parameter value A0013971 Escape...

Страница 63: ...and text editor Numeric editor symbols Symbol Meaning 0 9 A0013998 Selection of numbers from 0 to 9 A0016619 Inserts decimal separator at the input position A0016620 Inserts minus sign at the input po...

Страница 64: ...3986 Exits the input without applying the changes A0014040 Clears all entered characters Correction symbols under A0013989 Clears all entered characters A0013991 Moves the input position one position...

Страница 65: ...Calling up and closing the context menu The user is in the operational display 1 Press for 2 s The context menu opens XXXXXXXXXX kg h 20 50 Setup Conf backup disp Simulation A0014003 EN 2 Press simul...

Страница 66: ...HART 66 Endress Hauser 8 3 5 Envelope curve on the display and operating module In order to assess the measuring signal the envelope curve and if a mapping has been recorded the mapping curve can be...

Страница 67: ...2 HART device variables and measuring values On delivery the following measruing values are assigned to the HART device varaibles Device variables for interface measurements Device variable Measuring...

Страница 68: ...device appears 1 A0025866 1 Commissioning button calls up the wizard 3 Click on Commissioning to call up the wizard 4 Enter or select the appropriate value for each parameter These values are immediat...

Страница 69: ...oint Checklist Post installation check 34 Checklist Post connection check 47 11 2 Setting the operating language Factory setting English or ordered local language XXXX XXX X X 20 50 Operation Setup Ma...

Страница 70: ...nter tag for measuring point 2 Navigate to Setup Operating mode Select Interface with capacitance option 3 Navigate to Setup Distance unit Select distance unit 4 Navigate to Setup Tank type Select tan...

Страница 71: ...nk empty option NOTICE Wrong dielectric constant of the lower medium may cause a measuring error If in the case of Operating mode Interface with capacitance the lower medium is not water it is necessa...

Страница 72: ...rameter Navigation in the menu Expert Diagnostics Envelope diagnostics Save reference curve Meaning of the options No No action Yes The current envelope curve is saved as reference curve In devices wh...

Страница 73: ...put Factory setting for devices with 2 current outputs Format display 1 value max size 1 value max size Value 1 display Interface linearized Interface linearized Value 2 display Level linearized Level...

Страница 74: ...corresponding linearized value 100 or the corresponding linearized value 2 1 Level linearized 0 or the corresponding linearized value 100 or the corresponding linearized value 1 for devices with 2 cur...

Страница 75: ...evice Duplicate The transmitter configuration is duplicated to another device using the transmitter display module The following parameters which characterize the individual measuring point are not in...

Страница 76: ...FMP55 HART 76 Endress Hauser 11 8 Protection of the settings against unauthorized changes There are two ways to protect the settings against unauthorized changes Via parameter settings software lockin...

Страница 77: ...the plug correctly Display is defective Replace display Communication error is indicated on the display when starting the device or connecting the display Electromagnetic interference Check grounding...

Страница 78: ...13 If necessary select a more detailed setting in the Medium property parameter Level echo suppressed Delete mapping and record new mapping curve if required Record map parameter 120 Device displays a...

Страница 79: ...rating elements Status signals A0013956 Failure A device error is present The measured value is no longer valid A0013959 Function check The device is in service mode e g during a simulation A0013958 O...

Страница 80: ...A0013962 A0013958 441 Curr output 1 3 digit number If two or more diagnostic messages are pending simultaneously only the message with the highest priority is shown Additional pending diagnostic messa...

Страница 81: ...ior with diagnostic code 5 Operation time of occurrence 6 Remedial measures The user is in the diagnostic message 1 Press symbol Diagnostic list submenu opens 2 Select the desired diagnostic event wit...

Страница 82: ...cs parameter the diagnostic event is shown with event text 2 On the right in the display range hover the cursor over the Actual diagnostics parameter A tool tip with remedial measures for the diagnost...

Страница 83: ...dules 3 Change I O Modul or main electronics F Alarm 262 Module connection 1 Check module connections 2 Change electronic modules F Alarm 270 Main electronic failure Change main electronic module F Al...

Страница 84: ...tivate simulation switch output C Warning 495 Diagnostic event simulation Deactivate simulation C Warning 585 Simulation distance Deactivate simulation C Warning Diagnostic of process 801 Energy too l...

Страница 85: ...and closing the remedial measures 1 Press The message for the remedial measures for the selected diagnostic event opens 2 Press simultaneously The message about the remedial measures closes 12 6 2 Fil...

Страница 86: ...error event list I1184 Display connected I1185 Display backup done I1186 Restore via display done I1187 Settings downloaded with display I1188 Display data cleared I1189 Backup compared I1256 Display...

Страница 87: ...z Support of SD03 additional languages HistoROM functionality enhanced Advanced Diagnostic function block integrated Improvements and bugfixes BA01003F 00 EN 15 13 BA01003F 00 EN 16 14 GP01000F 00 EN...

Страница 88: ...ttack the surface of the hosuing and the seals 13 2 Cleaning coax probes For cleaning purposes the ground tube can be stripped of the coaxial probe When unmounting and remounting the tube make sure th...

Страница 89: ...parts Carry out repairs according to the instructions On completion of repairs carry out the specified routine test on the device Only Endress Hauser Service may convert a certified device into a dif...

Страница 90: ...al information www endress com deviceviewer 71023451 XPF0001 XPF0002 XPF0003 71023451 71023451 71023451 71023451 3 4 1 5 2 6 A0014979 23 Example for spare part nameplate in connection compartment cove...

Страница 91: ...5466 298 5 11 8 273 7 10 8 255 1 10 1 5 6 6 1 4 a 164 6 46 b 3 5 6 5 5 5 A0015472 24 Weather protection cover Dimensions mm in a 37 8 mm 1 5 in b 54 mm 2 1 in The weather protection cover can be order...

Страница 92: ...ering star Accessory Description Centering star PFA 16 4 mm 0 65 in 37 mm 1 46 in can be used for FMP55 10 0 39 A 16 4 0 65 B 37 1 46 A0014577 A For 8 mm 0 3 in probes B For 12 mm 0 47 in and 16 mm 0...

Страница 93: ...ected in feature 050 Option Measurement Device If a device has not been ordered in the version Prepared for display FHX50 but is nevertheless to be equipped with an FHX50 it is essential to select the...

Страница 94: ...s to retrofitted with the overvoltage protection Order code for retrofitting For 1 channel devices feature 020 option A OVP10 71128617 For 2 channel devices feature 020 option B C E or G OVP20 7112861...

Страница 95: ...devices to a WirelessHART network The WirelessHART adapter can be mounted directly at a HART device and is easly integrated into an existing HART network It ensures safe data transmission and can be o...

Страница 96: ...ponents Accessory Description Graphic Data Manager Memograph M The graphic data manager Memograph M provides information on all the relevant process variables Measured values are recorded correctly li...

Страница 97: ...on Operating menu Language 160 Setup 111 Device tag 111 Operating mode 111 Distance unit 111 Tank type 112 Tube diameter 112 DC value 116 Medium group 113 Empty calibration 113 Full calibration 114 Le...

Страница 98: ...rocess property 124 Interface property 124 DC value lower medium 125 Level unit 126 Blocking distance 126 Level correction 127 Automatic DC calculation 130 Manual thickness upper layer 130 DC value 13...

Страница 99: ...amp at echo lost 143 Blocking distance 143 SIL WHG confirmation 145 Deactivate SIL WHG 146 Reset write protection 146 Code incorrect 146 Probe settings 147 Probe grounded 147 Probe length correction 1...

Страница 100: ...it 155 Assign diagnostic behavior 155 Switch on value 156 Switch on delay 157 Switch off value 157 Switch off delay 158 Failure mode 158 Switch status 158 Invert output signal 158 Display 160 Language...

Страница 101: ...on management 166 Comparison result 167 Administration 169 Define access code 171 Define access code 171 Confirm access code 171 Device reset 169 Diagnostics 172 Actual diagnostics 172 Previous diagno...

Страница 102: ...ice type 178 Manufacturer ID 178 Measured values 179 Distance 115 Level linearized 136 Interface distance 118 Interface linearized 137 Thickness upper layer 180 Output current 1 to 2 153 Measured curr...

Страница 103: ...urrent output 1 to 2 simulation 188 Value current output 1 to 2 188 Switch output simulation 188 Switch status 189 Device alarm simulation 189 Device check 190 Start device check 190 Result device che...

Страница 104: ...111 Distance unit 111 Tank type 112 Tube diameter 112 Medium group 113 Empty calibration 113 Full calibration 114 Level 115 Distance 115 Signal quality 116 DC value 116 Interface 117 Interface distan...

Страница 105: ...orrection 127 Manual thickness upper layer 127 Measured thickness upper layer 128 DC value 128 Calculated DC value 128 Use calculated DC value 129 Linearization 132 Linearization type 134 Unit after l...

Страница 106: ...SIL WHG confirmation 145 Deactivate SIL WHG 146 Reset write protection 146 Code incorrect 146 Probe settings 147 Probe grounded 147 Present probe length 147 Confirm probe length 148 Current output 1...

Страница 107: ...57 Switch off value 157 Switch off delay 158 Failure mode 158 Switch status 158 Invert output signal 158 Display 160 Language 160 Format display 160 Value 1 to 4 display 162 Decimal places 1 to 4 162...

Страница 108: ...sult 167 Administration 169 Define access code 171 Device reset 169 Diagnostics 172 Actual diagnostics 172 Timestamp 172 Previous diagnostics 172 Timestamp 173 Operating time from restart 173 Operatin...

Страница 109: ...36 Interface distance 118 Interface linearized 137 Thickness upper layer 180 Output current 1 to 2 153 Measured current 1 181 Terminal voltage 1 181 Data logging 182 Assign channel 1 to 4 182 Logging...

Страница 110: ...P55 HART 110 Endress Hauser Switch status 189 Device alarm simulation 189 Device check 190 Start device check 190 Result device check 190 Last check time 190 Level signal 191 Launch signal 191 Interfa...

Страница 111: ...ting FMP5x Operating mode Navigation Setup Operating mode Prerequisite The device has the interface measurement application package available for FMP51 FMP52 FMP54 3 FMP55 always contains this package...

Страница 112: ...ditional information Depending on the probe some of the options mentioned above may not be available or there may be additional options For coax probes the default setting is Tank type Coaxial and can...

Страница 113: ...ameter as follows Medium group Medium property Others Unknown Water based DC 4 DC 4 7 The Medium property parameter can be changed at a later point of time However when doing so the Medium group param...

Страница 114: ...the interface level Full calibration Navigation Setup Full calibr Description Specify the distance F between the minimum level 0 and the maximum level 100 User entry Depending on the probe Factory se...

Страница 115: ...unit is defined in the Level unit parameter 126 In case of interface measurements this parameter always refers to the total level Distance Navigation Setup Distance Description Displays the measured...

Страница 116: ...ted echo either the level interface echo 4 or the end of probe echo To differentiate between these two the quality of the end of probe echo is always displayed in brackets In case of a lost echo Signa...

Страница 117: ...in various industries refer to the Endress Hauser DC manual CP01076F the Endress Hauser DC Values App available for Android and iOS Interface Navigation Setup Interface Prerequisite Operating mode 11...

Страница 118: ...interface Additional information DI DI A0013202 The unit is defined in the Distance unit parameter 111 Confirm distance Navigation Setup Confirm distance Description Specify whether the measured dista...

Страница 119: ...g the complete measuring range To be selected if the tank is completely empty The device records a mapping covering the complete measuring range minus Map gap to LN Factory map To be selected if the p...

Страница 120: ...arameter It indicates up to which distance a mapping has already been recorded Record map Navigation Setup Record map Prerequisite Confirm distance 118 Manual map or Distance too small Description Sta...

Страница 121: ...izard two parameters are displayed simultaneously on the display module at any one time The upper parameter can be edited whereas the lower parameter is displayed for reference purposes only Navigatio...

Страница 122: ...WHG mode is activated Writing access to the relevant parameters is denied Temporarily locked priority 4 Write access to the parameters is temporarily locked on account of internal processes in progres...

Страница 123: ...parameter 122 Enter access code Navigation Setup Advanced setup Ent access code Description Enter access code to disable write protection of parameters User entry 0 to 9999 Additional information For...

Страница 124: ...justs the signal evaluation filters and the damping of the output signal to the typical rate of level change defined in this parameter Process property Step response time s Fast 1 m 40 in min 5 Standa...

Страница 125: ...Capacitance Measurement Standard Condition The dielectric constant of the upper medium is known Signal evaluation As long as a clear interface is detected the specific capacitance pF m is continuousl...

Страница 126: ...in the Distance unit parameter is used for the basic calibration Empty calibration 113 and Full calibration 114 The unit defined in the Level unit parameter is used to display the unlinearized level...

Страница 127: ...Level correction Description Specify level correction if required User entry 200000 0 to 200000 0 Factory setting 0 0 Additional information The value specified in this parameter is added to the measu...

Страница 128: ...onstant of the upper medium Measured thickness upper layer Navigation Setup Advanced setup Interface Meas thick u lay Description Displays the measured interface thickness Thickness UP of the upper me...

Страница 129: ...used Selection Save and exit Cancel and exit Factory setting Cancel and exit Additional information Meaning of the options Save and exit The calculated constant is assumed to be the correct one Cancel...

Страница 130: ...two parameters are displayed simultaneously on the display module at any one time The upper parameter can be edited whereas the lower parameter is displayed for reference purposes only Navigation Set...

Страница 131: ...4 None B Linearization type 134 Linear C Linearization type 134 Table D Linearization type 134 Pyramid bottom E Linearization type 134 Conical bottom F Linearization type 134 Angled bottom G Lineariza...

Страница 132: ...u on the display module Navigation Setup Advanced setup Linearization Linearization Linearization type 134 Unit after linearization 135 Free text 136 Maximum value 137 Diameter 137 Intermediate height...

Страница 133: ...avigation Setup Advanced setup Linearization Linearization Linearization type 134 Unit after linearization 135 Free text 136 Level linearized 136 Interface linearized 137 Maximum value 137 Diameter 13...

Страница 134: ...inearization Lineariz type Description Select linearization type Selection None Linear Table Pyramid bottom Conical bottom Angled bottom Horizontal cylinder Sphere Factory setting None Additional info...

Страница 135: ...ht 138 The height of the pyramid Conical bottom The output value corresponds to the volume or weight in a tank with conical bottom The following additional parameters have to be specified Unit after l...

Страница 136: ...ation mode In order to define the new level unit select the Free text option in the Unit after linearization parameter and enter the required unit into the Free text parameter 136 Free text Navigation...

Страница 137: ...alue Prerequisite Linearization type 134 has one of the following values Linear Pyramid bottom Conical bottom Angled bottom Horizontal cylinder Sphere Description Specify the maximum content of the ve...

Страница 138: ...om Description Specify intermediate height H User entry 0 to 200 m Factory setting 0 m Additional information 0 H A0013264 H Intermediate height The unit is defined in the Distance unit parameter 111...

Страница 139: ...isting table is deleted and the complete table is entered again To do so delete the existing table Table mode 138 Clear table Then enter a new table How to enter the table Via FieldCare The table poin...

Страница 140: ...nearization type 134 Table Table mode 138 Semiautomatic Description Displays measured level value before linearization This value is transmitted to the table Customer value Navigation Setup Advanced s...

Страница 141: ...The measured level is not linearized If Linearization type 134 Table at the same time the device issues error message F435 Enable The measured level is linearized according to the table When editing t...

Страница 142: ...he output value is continously shifted towards 0 or 100 The slope of the ramp is defined in the Ramp at echo lost parameter 143 Value echo lost In the case of a lost echo the output assumes the value...

Страница 143: ...e measuring range per minute min For a negative slope of the ramp The measured value is continuously decreased until it reaches 0 For a positive slope of the ramp The measured value is continuosly inc...

Страница 144: ...me history or Long time history Expert Sensor Gas phase comp GPC mode On Without correction or External correction If one of these conditions is not met signals in the blocking distance will always be...

Страница 145: ...Approval option LA SIL or LC WHG overfill prevention which are currently not in the SIL or WHG locked state The SIL WHG confirmation wizard is required to lock the device according to SIL or WHG For...

Страница 146: ...tion Setup Advanced setup Deactiv SIL WHG Res write prot Description Enter unlocking code User entry 0 to 65535 Factory setting 0 Code incorrect Navigation Setup Advanced setup Deactiv SIL WHG Code in...

Страница 147: ...e the map using the Record map parameter 120 before performing the automatic probe length correction After the probe length correction a new map can be recorded using the Record map parameter 120 Alte...

Страница 148: ...ocedure has to be repeated until the displayed value matches the actual length of the probe Probe length too big To be selected if the displayed length is bigger than the actual length of the probe A...

Страница 149: ...tool all parameters concerning the probe length correction are located directly in the Probe settings submenu 147 Navigation Setup Advanced setup Probe settings Prob length corr Confirm probe length...

Страница 150: ...rface amplitude Factory setting For interface measurements Current output 1 Interface linearized Current output 2 9 Level linearized Additional information Definition of the current range for the proc...

Страница 151: ...dvanced setup Curr output 1 to 2 Current span Description Select current range for process variable and alarm signal Selection 4 20 mA 4 20 mA NAMUR 4 20 mA US Fixed current Factory setting 4 20 mA NA...

Страница 152: ...setting 0 0 s Additional information Fluctuations of the measured value affect the output current with an exponential delay the time constant of which is defined in this parameter With a small time c...

Страница 153: ...t follows the actual measured value the error is ignored Defined value The output current assumes the value defined in the Failure current parameter 153 The error behavior of other output channels is...

Страница 154: ...agnostic behavior parameter 155 determines for which type of event the output is opened Limit The output is normally closed and is only opened if a measured variable exceeds or falls below a defined l...

Страница 155: ...or limit monitoring Selection Off Level linearized Distance Interface linearized Interface distance Thickness upper layer Terminal voltage Electronic temperature Measured capacitance Relative echo amp...

Страница 156: ...relative position of the Switch on value and Switch off value parameters Switch on value Switch off value The output is closed if the measured value is larger than Switch on value The output is opened...

Страница 157: ...Off Description Define switch on delay User entry 0 0 to 100 0 s Factory setting 0 0 s Switch off value Navigation Setup Advanced setup Switch output Switch off value Prerequisite Switch output functi...

Страница 158: ...mode Navigation Setup Advanced setup Switch output Failure mode Description Define output behavior in alarm condition Selection Actual status Open Closed Factory setting Open Switch status Navigation...

Страница 159: ...Operating menu Endress Hauser 159 Additional information Meaning of the options No The behavior of the switch output is as described above Yes The states Open and Closed are inverted as compared to t...

Страница 160: ...s Portuguesa Polski Russian Svenska T rk e Chinese Japanese Korean Bahasa Indonesia ti ng Vi t Vietnamese e tina Czech Factory setting The language selected in feature 500 of the product structure If...

Страница 161: ...mm 23 5 V 1 1 1 V A0019966 36 Format display 1 value large 2 values 93 5 159 0 mm 93 5 V 26 3 C 1 1 1 1 V A0019968 37 Format display 4 values The Value 1 to 4 display 162 parameters specify which meas...

Страница 162: ...t Value 1 display Interface linearized Value 2 display Level linearized Value 3 display Thickness upper layer Value 4 display Current output 1 For interface measurements and two current outputs Value...

Страница 163: ...rmat can display simultaneously Display damping Navigation Setup Advanced setup Display Display damping Description Define display reaction time to fluctuations in the measured value User entry 0 0 to...

Страница 164: ...etup Display Separator Description Select decimal separator for displaying numerical values Selection Factory setting Number format Navigation Setup Advanced setup Display Number format Description Ch...

Страница 165: ...ith optical keys Description Switch the local display backlight on and off Selection Disable Enable Factory setting Disable Additional information Meaning of the options Disable Switches the backlight...

Страница 166: ...the Operating mode parameter 111 Navigation Setup Advanced setup Conf backup disp Operating time Navigation Setup Advanced setup Conf backup disp Operating time Description Indicates how long the devi...

Страница 167: ...mpared to the current device configuration of the HistoROM The result of this comparison is displayed in the Comparison result parameter 167 Clear backup data The backup copy of the device configurati...

Страница 168: ...of the HistoROM is corrupt or not compatible with the backup copy in the display module Check not done The device configuration of the HistoROM has not yet been compared to the backup copy in the dis...

Страница 169: ...all parameters marked with the symbol in this document On the local display the symbol in front of a parameter indicates that the parameter is write protected Once the access code has been defined wri...

Страница 170: ...been ordered This option is only visible if customer specific settings have been ordered Of customer settings All customer parameters are reset to their factory setting Service parameters however rem...

Страница 171: ...Administration submenu The Confirm access code parameter is not available for operation via operating tool Navigation Setup Advanced setup Administration Def access code Define access code Navigation...

Страница 172: ...essage and remedy measures can be viewed via the symbol on the display Timestamp Navigation Diagnostics Timestamp Description Displays timestamp for the Actual diagnostics parameter 172 User interface...

Страница 173: ...rating time from restart Navigation Diagnostics Time fr restart Description Displays the time the device has been in operation since the last device restart User interface Days d hours h minutes m sec...

Страница 174: ...Display the current diagnostics messages with the highest to fifth highest priority Additional information The display consists of Symbol for event behavior Code for diagnostic behavior Operating time...

Страница 175: ...ter is only used for operation via the local display The status signals are categorized according to NAMUR NE 107 Event list submenu The Event list submenu displays the history of past events of the c...

Страница 176: ...number To identify the device quickly e g when contacting Endress Hauser To obtain specific information on the device using the Device Viewer www endress com deviceviewer The serial number is also ind...

Страница 177: ...Additional information The extended order code indicates the version of all the features of the product structure and thus uniquely identifies the device Device revision Navigation Diagnostics Device...

Страница 178: ...he device is registered with the the HART Communication Foundation Additional information The device type is needed to allocate the correct Device Description file DD to the device Manufacturer ID Nav...

Страница 179: ...ation DL DL A0013199 38 Distance for interface measurements The unit is defined in the Distance unit parameter 111 Level linearized Navigation Diagnostics Measured val Level linearized Description Dis...

Страница 180: ...terface linearized Navigation Diagnostics Measured val Interf lineariz Prerequisite Operating mode 111 Interface or Interface with capacitance Description Displays the linearized interface height Addi...

Страница 181: ...igation Diagnostics Measured val Output curr 1 to 2 Description Displays calculated output current Measured current 1 Navigation Diagnostics Measured val Measur curr 1 Prerequisite Only available for...

Страница 182: ...plitude Relative interface amplitude Absolute EOP amplitude EOP shift Noise of signal Calculated DC value Analog output adv diagnostics 1 Analog output adv diagnostics 2 Factory setting Off Additional...

Страница 183: ...t log If 3 logging channels are used T log 333 t log If 4 logging channels are used T log 250 t log Once this time elapses the oldest data points in the data log are cyclically overwritten such that a...

Страница 184: ...function Event List HistoROM The Display channel 1 to 4 submenus invoke a diagram of the logging history of the respective channel x axis depending on the number of selected channels 250 to 1000 meas...

Страница 185: ...which can be simulated Condition to be simulated Associated parameters Specific value of a process variable Assign measurement variable 187 Process variable value 187 Specific value of the output cur...

Страница 186: ...igation Expert Diagnostics Simulation Simulation Assign measurement variable 187 Process variable value 187 Current output 1 to 2 simulation 188 Value current output 1 to 2 188 Switch output simulatio...

Страница 187: ...e Process variable value parameter 187 If Assign measurement variable Off a simulation is active This is indicated by a diagnotic message of the Function check C category Process variable value Naviga...

Страница 188: ...ation Expert Diagnostics Simulation Value curr out 1 to 2 Prerequisite Current output simulation 188 On Description Enter current value for the simulation User entry 3 59 to 22 5 mA Factory setting 3...

Страница 189: ...larm simulation on or off Selection Off On Factory setting Off Additional information When selecting the On option the device generates an alarm This helps to check the correct output behavior of the...

Страница 190: ...f the device check Additional information Meaning of the display options Installation ok Measurement possible without restrictions Accuracy reduced A measurement is possible However the measuring accu...

Страница 191: ...ics Device check Launch signal Prerequisite Device check has been performed Description Displays result of the display check for the launch signal User interface Check not done Check not OK Check OK A...

Страница 192: ...eartbeat submenu The Heartbeat submenu is only available via FieldCare or DeviceCare It contains the wizards which are part of the Heartbeat Verification and Heartbeat Monitoring application packages...

Страница 193: ...t 1 to 2 Submenu 150 Current output 1 to 2 simulation Parameter 188 Current span Parameter 151 Customer value Parameter 140 D Damping output Parameter 152 Data logging Submenu 182 DC value Parameter 1...

Страница 194: ...Interface property Parameter 124 Interface signal Parameter 191 Intermediate height Parameter 138 Invert output signal Parameter 158 K Keypad lock Disabling 58 Switch on 58 L Language Parameter 160 La...

Страница 195: ...tion Wizard 145 Simulation Submenu 186 187 Spare parts 90 Nameplate 90 Start device check Parameter 190 Status signals 60 79 Stilling well 26 Submenu Administration 169 Advanced setup 122 Configuratio...

Страница 196: ...put 1 to 2 Parameter 188 Value echo lost Parameter 142 W W M Device Viewer 90 Wizard Automatic DC calculation 130 Deactivate SIL WHG 146 Define access code 171 Mapping 121 Probe length correction 149...

Страница 197: ......

Страница 198: ...www addresses endress com 71344071 71344071...

Отзывы: