Operating menu
Levelflex FMP55 HART
116
Hauser
Signal quality
Navigation
Setup → Signal quality
Description
Displays the signal quality of the evaluated echo.
Additional information
Meaning of the display options
•
Strong
The evaluated echo exceeds the threshold by at least 10 mV.
•
Medium
The evaluated echo exceeds the threshold by at least 5 mV.
•
Weak
The evaluated echo exceeds the threshold by less than 5 mV.
•
No signal
The device does not find a usable echo.
The signal quality indicated in this parameter always refers to the currently evaluated
echo: either the level/interface echo
or the end-of-probe echo. To differentiate between
these two, the quality of the end-of-probe echo is always displayed in brackets.
In case of a lost echo (
Signal quality = No signal
) the device generates the following
error message:
• F941, for
Output echo lost (→ 142) = Alarm
.
• S941, if another option has been selected in
.
DC value
Navigation
Setup → DC value
Prerequisite
The device has the "interface measurement" application package
Description
Specify relative dielectric constant ε
r
of the upper medium (DC
1
).
User entry
1.0 to 100
Factory setting
2.0
4)
Of these two echos the one with the lower quality is indicated.
5)
Product structure: Feature 540 "Application Package", Option EB "Interface measurement"