Operating menu
Levelflex FMP55 HART
124
Hauser
"Interface" submenu
Navigation
Setup → Advanced setup → Interface
Process property
Navigation
Setup → Advanced setup → Interface → Process property
Description
Specify typical rate of change for the interface position.
Selection
• Fast > 1 m (40 in)/min
• Standard < 1 m (40in) /min
• Medium < 10 cm (4in) /min
• Slow < 1 cm (0.4in) /min
• No filter / test
Factory setting
Standard < 1 m (40in) /min
Additional information
The device adjusts the signal evaluation filters and the damping of the output signal to the
typical rate of level change defined in this parameter:
Process property
Step response time / s
Fast > 1 m (40 in)/min
5
Standard < 1 m (40in) /min
15
Medium < 10 cm (4in) /min
40
Slow < 1 cm (0.4in) /min
74
No filter / test
2.2
Interface property
Navigation
Setup → Advanced setup → Interface → Interface prop.
Prerequisite
Operating mode (→ 111)
Interface with capacitance
Description
Select interface property.
The interface property determines how the Guided Radar and the Capacitance
Measurement interact.
Selection
• Special: automatic DC
• Build up
• Standard
• Emulsion layer
Factory setting
Standard