![Endress+Hauser iTHERM TrustSens TM371 Скачать руководство пользователя страница 84](http://html1.mh-extra.com/html/endress-hauser/itherm-trustsens-tm371/itherm-trustsens-tm371_operating-instructions-manual_2408482084.webp)
Operating menu and parameter description
iTHERM TrustSens TM371, TM372
84
Hauser
Operating time
Navigation
Diagnostics → Operating time
Description
Use this function to display the length of time the device has been in operation up to now.
Display
Hours (h)
14.3.1 "Diagnostic list" submenu
Up to 3 diagnostic messages currently pending are displayed in this submenu. If more than
3 messages are pending, the messages with the highest priority are shown on the display.
Overview of all the diagnostic messages and remedies → 37.
Actual diagnostics count
Navigation
Diagnostics → Diagnostic list → Actual diagnostics count
Description
Use this function to display the number of diagnostic messages currently pending in the
device.
Actual diagnostics
Navigation
Diagnostics → Diagnostics list → Actual diagnostics
Description
Use this function to display the current diagnostic messages with the highest priority to
the third-highest priority.
Additional information
Example for display format:
F001-Device failure
Actual diag channel
Navigation
Diagnostics → Diagnostic list → Actual diag channel
Description
Indication of the sensor input which this diagnostic message refers.
Use this function to display the current diagnostic message. If two or more messages occur
simultaneously, the message with the highest priority is shown.