5.4.6
Device status and error messages
You can read the extended device status, in this case, current error messages, via Command "48".
The command delivers information which is partly coded in bits (see table below).
!
Note!
• For a detailed explanation of the device status and error messages and their elimination,
see
• Nothing is assigned to bits and bytes that are not listed.
Byte
Bit
Error No.
Short error description
0
0
001
Serious device error
1
011
Measuring amplifier has faulty EEPROM
2
012
Error when accessing data of the measuring amplifier EEPROM
1
1
031
S-DAT: defective or missing
2
032
S-DAT: Error accessing saved values
5
051
I/O and the amplifier are not compatible
3
3
111
Totalizer checksum error
4
121
I/O board and the amplifier board are not compatible
4
3
251
Internal communication error on the amplifier board
4
261
No data reception between amplifier and I/O board
5
0
321
Coil current of the sensor is outside the tolerance
7
339
Current buffer:
The buffering of the flow components (measuring mode with pulsating flow) could not
be calculated and output within 60 seconds.
6
0
340
1
341
2
342
3
343
Frequency buffer:
The buffering of the flow components (measuring mode with pulsating flow) could not
be calculated and output within 60 seconds.
4
344
5
345
6
346
7
347
Pulse buffer:
The buffering of the flow components (measuring mode with pulsating flow) could not
be calculated and output within 60 seconds.
7
0
348
1
349
2
350
3
351
Current output:
Flow is out of range.
4
352
5
353
6
354
7
355
Frequency output:
Flow is out of range.
8
0
356
1
357
2
358