General information
The clamping device must be installed in a way such that it can support the weight of the source
container and the Gammapilot M FMG60 under all expected operating conditions (e.g. vibrations).
If necessary, the customer should provide additional support with a separate stable, low-vibration
construction. Note weights: Gammapilot M FMG60: 14 to 29 kg (30.87 to 63.95 lbs)
Source container FQG60: max. 18 kg (39.69 lbs)
NOTICE
Mounting instructions are provided in the documentation:
SD00330F/00 and SD00331F/00.
Torque for the mounting
screws (supplied by
customer)
Post-installation check
Measuring the local dose rate
The local dose rate in the vicinity of the source container and the detector must be measured after
mounting the unit.
CAUTION
!
Depending on the installation, radiation can also occur outside the actual beam-emitting channel
through scattering.
‣
In such cases it must be screened by the use of additional lead or steel shielding.
‣
Render or mark all control and exclusion areas as prohibited for unauthorized entry.
Behavior in event of empty process vessel or pipe
CAUTION
!
Radiation
‣
Once the unit has been correctly mounted, the control area of the empty process vessel has to be
measured.
‣
If it is necessary, this area must be blocked off and marked. If there is an entry into the interior
space of the process vessel, it has to be closed and marked with a sign "radioactive".
‣
The entry is only allowed after checking all safety regulation by the responsible radiation
protection officer.
‣
If maintenance operations are carried out in or at the product vessel, it is mandatory to switch the
radiation OFF.
If the pipe becomes empty as a result of operational processes, the level of radiation on the detector
side can reach dangerous levels.
• In such instances, the radiation emission channel must be closed immediately for reasons of
radiation protection.
• A high local dose rate also causes the detector unit (scintillator and photomultiplier) to age quickly.
The best way to avoid such a situation is to mount a second radiometric measuring system that
monitors the radiation intensity. If high radiation levels occur, an alarm is issued and the source
container shall be switched OFF.
Material
Min. tensile strength
Coefficient of friction (
) Torque
Stainless steel
700 N/mm² (157.36 lbf)
0.14
32 Nm (23.6 lbf ft)