Orientation for density measurement on vertical pipes
If possible, density should be measured with a flow direction from bottom to top. With this type of
measuring arrangement, the Gammapilot M FMG60 should preferably be positioned with the terminal
head at the top. If this arrangement is not possible, an additional bracket must be used to secure the
Gammapilot M FMG60 against slipping.
A0018508
A
Feature 240 "Emission Angle; Application", Option model 3 "20 deg; limit density"
B
Feature 240 "Emission Angle; Application", Option model 4 "20 deg; density 30 deg diagonal radiation"
1
FMG60
Orientation for density measurement on horizontal pipes
With this type of orientation, it is advisable to mount the FQG60 above the pipe. This prevents the
buildup of solid particles or liquids on the shutter. However, attention should be paid to the effect of
air bubbles and material buildup in the pipe.
A0018509
A
Feature 240 "Emission Angle; Application", Option model 3 "20 deg; limit density"
B
Feature 240 "Emission Angle; Application", Option model 4 "20 deg; density 30 deg diagonal radiation"
1
FMG60
Lateral installation (see graphic) is only permitted in low-vibration applications, while taking safety
instructions into account (periodical inspection of the shutter, padlock or snap hook and mounting
clamps). A clamping device is available as an accessory for mounting the device on pipes (
29).