Source container FQG60
Hauser
23
After recalibration, perform the steps above in the reverse order. To bring the calibration plate to its
rest position, slide it to the lower limit stop.
To quickly check the density measurement, always recreate these constant conditions and check the
value displayed. If the values deviate, perform a recalibration (
30, associated documentation
"Gammapilot M FMG60").
The adjustment point "10" is available with the Gammapilot M device for recalibration. This point can
be entered if the measuring conditions have changed, for example due to buildup in the measuring
tube.
I
o
corresponds to the pulse rate when the pipe is empty. The value can be significantly larger than all
the pulse rates that actually occur during measurement. Once the information has been entered, I
o
is
recalculated to suit the current measuring conditions. The absorption coefficient
μ
is retained from the
original calibration.
WARNING
!
The calibration plate does not represent a shielding in the sense of radiation protection.