8 Function group "output" (06), - "profibus param." (06), PROFIBUS-PA only
Hauser
51
8.13
Function "output current" (067), HART only
Displays the output current in mA.
Caution!
This function is available for HART devices only!
8.14 Function "2nd cyclic value" (067),
PROFIBUS-PA only
Selects the second cyclical value.
•
height/dist.
The Prosonic M always transmits the distance as the second cyclical value.
Caution!
This function is available for PROFIBUS-PA devices only!
8.15 Function "4mA-value" (068), HART only
In this function specify the level (or volume, weight, flow resp.), at which the output cur-
rent should be 4 mA. This value will be used if you choose the option "curr. turn down"
in the
"current output mode" (063)
function.