Installation
Flowfit CUA252
12
Hauser
3.3.3
Installation with bubble trap
a0022693
Fig. 6:
Connection example with bubble trap
1
Supply from below
2
Shutoff valve
3
Bubble trap
4
Bubble trap venting (included in the delivery)
5
Shutoff valve (throttle to increase pressure)
6
Outlet
7
Transition to D 12 with connection for vent line (included in the delivery)
8
CUA252 flow assembly
9
CUS52D turbidity sensor
10
Transition to D 12
The lost water resulting from the bubble trap is not suitable for feedback into the
process.
For the hose system, use PVC hoses with an internal diameter of 12 mm (3/8").
Secure the hose system with worm drive hose clips (not included in the delivery).
The inlet and outlet connection of the flow assembly are always identical, the system is
symmetrical.
Install the flow assembly vertically.
The supply must be connected at the bottom (upward flow in the pipe).
Insert a orifice plate into the upper process connection (
å
6, item 7) of the flow
assembly to get the desired volume flow rate (included in the delivery):
• 1 mm for volume flow rate < 60 l/h (15.8 gal/hr)
• 3 mm for volume flow rate 60 to 100 l/h (15.8 to 26.4 gal/hr)
• 5 mm for volume flow rate > 100 l/h (26.4 gal/hr)
Avoid buckles and loops in the hose system.