Flowfit CUA252
Installation
Hauser
13
Pay attention to the installation instructions (flow direction) for the sensor.
Observe the maximum pressure and maximum temperature when operating the bubble
trap.
3.4
Sensor installation
Install the sensor so that the optical windows of the sensor (
å
8, item 1) are aligned
against the flow direction (item 2). The installation marking (item 3) helps you ensure the
correct sensor orientation.
3.5
Post-installation check
‣
After installation, check all the connections to ensure they are secure and leak-tight.
‣
Is the orientation correct?
‣
Make sure that the hoses can only be removed by force.
‣
Check all hoses for damage.
a0022864
Fig. 7: Sensor
installation
10
CUA252 flow assembly
11
Clamp seal
12
Locking clamp
13
CUS52D turbidity sensor
a0022863
Fig. 8: Sensor
orientation
1
Optical windows
2
Flow direction
3
Installation marking