Installation
Flexdip CYA112
18
Hauser
5. Above the cross clamp, mount the multifunctional clamp ring on the immersion tube
(funnel-shaped side facing upwards).
The multifunctional clamp ring acts as an anti-slip lock.
For detailed information on the CYH112 holder, see Operating Instructions
BA00430C
Installing the assembly with a float:
1. Mount the sensor → 13.
2. Mount the float on the CYH112 pendulum holder.
3. Align the assembly and holder.
4. Tighten the clamp screws finger-tight (finger-tight corresponds to 13 Nm (9.6 lbf ft).
5.2.8
Installing the quick fastener
1
2
3
4
A0016947
11
Quick fastener
1
Bore hole (makes it easier to screw the adapter on tightly)
2
Adapter
3
Union nut
4
O-rings
Installing the quick fastener:
1. Provide both O-rings (item 4) with a thin film of lubricant.
2. Screw the adapter (item 2) into the connection pipe of the immersion tube.
3. Insert a screwdriver or a similar tool (serves as lever) through the bores (item 1).
4. Tighten the adapter with the screwdriver.
5. Slide the union nut (item 3) over the adapter until the union nut engages with a click.
6. Route the sensor cable through the quick fastener and the immersion tube.
The Memosens coupling is secured in the quick fastener and cannot slip.
7. Connect the cable to the sensor.
8. Guide the sensor through the quick fastener (hold the sensor steady when doing so).
9. Using the quick fastener, install the sensor without twisting the measuring cable.
10. Turn the union nut to tighten the sensor.
11. Align the sensor if necessary.
Содержание Flexdip CYA112
Страница 2: ......
Страница 32: ...Accessories Flexdip CYA112 32 Endress Hauser A0035835 31 Example of mounted spray cleaning ...
Страница 36: ...www addresses endress com 71460314 71460314 ...