Installation
Flexdip CYA112
14
Hauser
6. Press the rubber plug onto the immersion tube.
7. Route the sensor cable downwards in a loop (do not bend).
8. Fit the splash protection cap.
The splash protection cap is fixed in place by the lip of the rubber plug.
Mounting sensors with a fixed cable
1. Route the sensor cable through the immersion tube.
2. Screw the sensor into the adapter or the connecting pipe.
3. In doing so, turn the assembly and not the sensor to ensure that the sensor cable is
not twisted.
4. Shorten the tip of the rubber plug to match the cable diameter.
5. Fit the rubber plug onto the sensor cable.
6. Press the rubber plug onto the immersion tube.
7. Route the sensor cable downwards in a loop (do not bend).
8. Fit the splash protection cap.
The splash protection cap is fixed in place by the lip of the rubber plug.
5.2.5
Fixed installation of the assembly on a transverse pipe
Prerequisite
The sensor is already mounted.
1
2
3
4
A0011292
5
Mounting the cross clamp
1
Main pipe
2
Cross clamp, closed side facing center of basin
3
Cross clamp, closed side facing upwards
4
Transverse pipe of CYH112 holder
‣
Mount the cross clamp in such a way that one closed side faces the center of the basin