6 Diagnosis and Troubleshooting
6.1 Alarms
Output on alarm
The following response is set in the factory:
If the FEC 12 detects a fault, then the current output is set to 22 mA (=110 %) to
enable process control systems to, e.g. assume a response.
Alarm indication
If no alarm indication is to be given on an alarm, then "
CONTINUE"
can be set. In this
case the FEC 12 continues to measure although the measured value is possibly
incorrect.
Step
Entry
Cursor in Display to
Significance
1 1
CALIBRATION
Basic setting
2
8
SAFETY ALARM
Output on alarm
MAX (110%)
Output jumps to 22 mA (=110 %), factory set
3
CONTINUE
Continue measuring, no alarm indication
4 F4
[ENTER]
Confirms entry
5
F2
[SEND]
Value to be transmitted
6.2 Simulation
Simulation enables the current value coming from the electronic insert to be
simulated. It can be used for checking the correct processing of probe signals, e.g. in
a process control system. Simulation may also be necessary for tracing faults. No
change in level is required to produce a different current value. The various types of
simulation
•
Level (
LEVEL)
•
Volume (
VOLUME)
•
Current (
CURRENT)
are independent from one another and therefore none affects the others.
Note!
•
If simulation is activated, Warning E 613 is displayed to indicate this.
•
The full simulation range is only available with a free probe, only a limited range is
possible with a covered probe.
•
On completion, switch off simulation again in order to continue normal measurement.
Simulation
Activate simulation as follows:
Step
Entry
Cursor in Display at
Significance
1 3
SERVICE/SIMULATION
Service/Simulation
2
5
SIMULATION
Simulation
OFF
Simulation switched off
3
ON
Simulation switched on
4 F4
[ENTER]
Confirms entry, value to be transmitted
Note!
Electronic insert FEC 12 (HART)
6 Diagnosis and Troubleshooting
Hauser
15