Condumax CLS15/16/21
Technical data
Hauser
21
9
Technical data
9.1
Input
9.1.1
Measured values
• Conductivity
• Temperature
9.1.2
Measuring ranges
Conductivity
(in relation to water at 25 °C (77 °F))
CLS15-A
0.04 to 20 μS/cm
CLS15-B/L
0.10 to 200 μS/cm
CLS16
0.04 to 500 μS/cm
CLS21
10 μS/cm to 20 mS/cm
Temperature
CLS15
-20 to 140 °C (-4 to 280 °F)
CLS16
-5 to 150 °C (23 to 300 °F)
CLS21
-20 to 135 °C (-4 to 275 °F)
9.1.3
Cell constant
CLS15-A
k = 0.01 cm
–1
CLS15-B/L
k = 0.1 cm
–1
CLS16
k = 0.1 cm
–1
CLS21
k = 1.0 cm
–1
, nominal
9.1.4
Temperature compensation
Pt 100 (class A as per IEC 60751) (CLS15) (CLS16) (CLS21)