5. Readily soluble biological buildup:
Rinse with pressurized water.
After cleaning or regeneration, you must rinse the sensor thoroughly with water and
then recalibrate it.
8
Repair
8.1
Sealing ring replacement and recalibration (CLS16 only)
Intact seals are a prerequisite for safe and reliable measurements. The seal should be replaced
at regular intervals to guarantee maximum sensor operational safety and hygiene.
Practical repair intervals can only be determined by the user as they depend greatly on the
operating conditions, such as:
• Type and temperature of the product
• Type and temperature of the cleaning agent
• Number of cleanings
• Number of sterilizations
• Operating environment
Recommended intervals for seal replacement (reference values)
Application
Window
Media with temperatures from 50 to 100 °C (122 to 212 °F)
Approx. 18 months
Media with temperatures < 50 °C (122 °F)
Approx. 36 months
Sterilization cycles, max. 150 °C (302 °F), 45 min.
Approx. 400 cycles
To ensure your sensor is operational again after being exposed to very high loads, you can
have it regenerated in the factory. In the factory, the sensor is fitted with new seals and
recalibrated.
Please contact your sales office for information on replacing the seal and recalibration in the
factory.
8.2
Return
The product must be returned if repairs or a factory calibration are required, or if the wrong
product was ordered or delivered. As an ISO-certified company and also due to legal
regulations, Hauser is obliged to follow certain procedures when handling any
returned products that have been in contact with medium.
To ensure swift, safe and professional device returns, please read the return procedures and
conditions at
www.endress.com/support/return-material
.