Technical data
Conducal CLY421
20
Hauser
7
Technical data
7.1
Input
Measured variables
μ
S/cm or M
cm; configurable
7.2
Power supply
Supply voltage
Wide-area power pack 100 to 240 VAC, 47 to 63 Hz, Class II equipment with functional
grounding
Rechargeable battery
Integrated lithium ion battery 14.4 V; 2.4 Ah
The fully charged battery enables a calibration kit operating time of over 80 hours.
External sensor cable
connection
Buccaneer plug, 6-pin , IP 68
7.3
Performance characteristics
Error calculation
Reference system adjustment with standard NIST reference material
Conducal adjustment with 5
μ
S/cm (or 200 k
cm)
The change of the CLS15D cell constant in the conductivity range between the standard
reference material and 5
μ
S/cm is not considered.
Depending on availability, different standard NIST reference materials are used to
adjust the reference system. Normally SRM 3191 is used with a nominal value of 100
μ
S/cm.
Reference devices
Uncertainty of reference solution
0.2 %
Uncertainty of temperature measurement
<< 0.1 %
Uncertainty of reference system display
0.2 %
Total uncertainty of reference system adjustment
0.3 %
Uncertainty of reference system adjustment
0.3 %
Uncertainty of reference system measurement at 5
μ
S/cm
0.6 %
Uncertainty of Conducal display at 5
μ
S/cm
0.6 %
Total uncertainty of Conducal adjustment with 5
μ
S/cm
0.9 %
(Only corresponds to the uncertainty of Conducal. The adjustment of measuring points with
Conducal requires an additional uncertainty analysis)
Reference measuring device used
Liquiline CM42
Reference measuring cell used
Condumax CLS15D