Conducal CLY421

Device description and operation

Hauser

13

Preparatory steps for comparison measurement with a bypass

 

Install the measuring arrangement as follows:

1.

Secure the flow assembly with the carrier on a pipe (e.g. railing). 
On square pipes mount the clamping jaw with the V-notch facing outwards, and facing 
inwards on circular pipes, 
or position the flow assembly safely.

2.

Set the "Bypass - Direct" switch to "Bypass" (Position 1).

3.

Using a hose connection adapter (provided in the case) mount the outflow hose on the 
"OUT" outlet of the flow assembly (Item 7). Only screw the hose connection adapter 
finger-tight into the flow assembly. 

4.

Using a hose connection adapter mount the medium hose on the "IN 1" inlet of the flow 
assembly (Item 4).

5.

Seal the "IN 2" inlet (Item 5) with a stopper (provided in the case).

6.

Screw the sensor of the calibration kit into the flow assembly (Item 2).

7.

In the flow assembly, seal the installation point for the process sensor (Item 3) with a 
stopper.

Preparatory steps for direct comparison measurement 

Install the measuring arrangement as follows:

1.

Secure the flow assembly with the carrier on a pipe (e.g. railing). 
On square pipes mount the clamping jaw with the V-notch facing outwards, and facing 
inwards on circular pipes, 
or position the flow assembly safely.

2.

Set the "Bypass - Direct" switch to "Direct" (Position 2).

3.

Using a hose connection adapter (provided in the case) mount the outflow hose on the 
"OUT" outlet of the flow assembly (Item 7). Only screw the hose connection adapter 
finger-tight into the flow assembly. 

4.

Using a hose connection adapter mount the medium hose on the "IN 2" inlet of the flow 
assembly (Item 5).

5.

Seal the "IN 1" inlet (Item 4) with a stopper (provided in the case).

6.

Screw the sensor of the calibration kit into the flow assembly (Item 2).

7.

Screw the process sensor into the flow assembly (Item 3). For sensors with a G1 process 
connection, use the G1 adapter clamp (provided in the case).

3.2.5

Electrical connection

Connecting the calibration kit:

1.

Install the measuring cable between the sensor, calibration kit and transmitter (exterior 
of the case).

2.

In the case of direct comparison measurement: 
Install the measuring cable between the process sensor and process transmitter.

3.

If a power supply is available, connect the mains cable (exterior of the case).

Содержание Conducal CLY421

Страница 1: ...Products Solutions Services BA00496C 07 EN 01 12 71142937 Operating Instructions Conducal CLY421 Conductivity calibration kit for ultrapure water applications...

Страница 2: ...Endress Hauser...

Страница 3: ...ery 6 2 3 Certificates and approvals 6 3 Device description and operation 7 3 1 Calibration kit design 7 3 2 Operation 8 4 Maintenance 15 4 1 Cleaning 15 4 2 Recalibration 15 4 3 Repair 15 4 4 Return...

Страница 4: ...measuring point check that all the connections are correct Ensure that electrical cables and hose connections are not damaged Do not operate damaged products Secure them against unintentional commissi...

Страница 5: ...non compliance if applicable Corrective action This symbol alerts you to a dangerous situation Failure to avoid the situation can result in a fatal or serious injury CAUTION Causes consequences Conse...

Страница 6: ...d approvals Declaration of Conformity The product meets the requirements of the harmonized European standards As such it complies with the legal specifications of the EC directives The manufacturer co...

Страница 7: ...on for measuring cable with cap 2 Connection for power supply with cap 1 Charger 7 G1 adapter clamp 2 Connections for power supply and measuring cable 8 Compartment for accessories 3 On Off switch for...

Страница 8: ...s being charged Green The battery is fully charged It can take several hours to charge the battery 3 2 2 Operating concept of the transmitter a0010027 Fig 3 Press the soft key select the menu directly...

Страница 9: ...h on temperature compensation if the difference in temperature between the calibration sensor and the process sensor is 0 1 C This is described in the Comparison measurement section 3 2 3 Measuring ar...

Страница 10: ...ison measurement point are the same This is ensured by Using short hose connections Wait until the temperature in the flow assembly adjusts to match the process temperature a0014022 Fig 9 Measuring ar...

Страница 11: ...case of direct comparison measurement all the important parameters match Temperature and Absolutely identical medium As it is necessary to remove the sensor from the process the medium can be contami...

Страница 12: ...nt with a bypass Direct comparison measurement a0014030 Fig 11 Flow assembly 1 2 3 4 5 6 7 8 1 Switchover from bypass 1 to direct 2 5 Inlet for direct comparison measurement 2 Installation point for c...

Страница 13: ...ring arrangement as follows 1 Secure the flow assembly with the carrier on a pipe e g railing On square pipes mount the clamping jaw with the V notch facing outwards and facing inwards on circular pip...

Страница 14: ...temperature difference is 0 1 C you do not need to change any settings on the transmitter of the calibration kit If the temperature difference is 0 1 C set the temperature compensation on both transm...

Страница 15: ...g on the operating conditions and the frequency of use the conductivity calibration kit must be calibrated regularly at the factory annual recalibration is recommended You then receive an updated fact...

Страница 16: ...ress Hauser 5 Accessories Recalibration Depending on the operating conditions and the frequency of use the conductivity calibration kit must be calibrated regularly at the manufacturer s factory Recom...

Страница 17: ...tter in front of the error number F Failure general error message M Maintenance required Action must be taken where applicable the measured value is still valid C Device is in service Check no error U...

Страница 18: ...tion to finish 131 PV not stable M Sensor too old Defective cable or connector 132 Temperature not stable M 133 Polarization warning M 180 Calibration expired alarm M 183 Usage 80 C warning M 194 Usag...

Страница 19: ...n 841 PV lower limit M 842 Temp upper limit M 843 Temp lower limit M 950 Conc temp lower limit M Check your process conditions Adjust the measuring range where necessary These messages only apply for...

Страница 20: ...ge between the standard reference material and 5 S cm is not considered Depending on availability different standard NIST reference materials are used to adjust the reference system Normally SRM 3191...

Страница 21: ...closed case without power cord Use indoors pollution level II 7 5 Process Process temperature 0 to 100 C 32 to 210 F Process pressure Max 6 bar 87 psi Minimum flow 30 l h 8 gal hr 7 6 Mechanical const...

Страница 22: ...17 Disposal 15 E Electrical connections 7 F Flow assembly 12 I Icons Safety conventions 4 Installation 4 12 M Maintenance 15 Measuring arrangements 9 N Nameplate 6 Notes on safety conventions and icon...

Страница 23: ...Conducal CLY421 Endress Hauser 23...

Страница 24: ...www addresses endress com 71142937...

Отзывы: