Product Structure
PMP 48
Cerabar M PMP 48
Certificates
R
Standard
G ATEX II 1/2 G, EEx ia IIC T6
H
ATEX II 2 G, EEx ia IIC T6
N
ATEX II 3 G, EEx nA II T5
K
ATEX II 1/2 D, EEx ia IIC T6
L
ATEX II 1/3 D (non-Ex power supply)
C
CSA General Purpose
S
CSA IS (suitable for Div. 2) Cl. I, II, III, Div. 1, Groups A…G
T
CSA Cl. I, Div. 2, Groups A…D, Cl. II, III, Div. 1, Groups E…G
P
FM IS (non-incendive) Cl. I, II, III, Div. 1, Groups A…G
M FM DIP, Cl. II, III, Div. 1, Groups E…G
V
TIIS Ex ia IIC T6
Housing
Stainless steel 1.4404 (AISI 316L)
Aluminium
E1
Cable gland M 20x1.5
E2
Cable gland M 20x1.5
G1
Cable entry G ½
G2
Cable entry G ½
C1
Cable entry ½ NPT
C2
Cable entry ½ NPT
H1
Harting plug (Han7D)
H2
Harting plug (Han7D)
L1
Plug M 12x1
L2
Plug M 12x1
K1
5 m cable (with pressure compensation)
K2
5 m cable (with pressure compensation)
Metal Sensor: Nominal Value (Maximum Overload)
Gauge pressure
3H 0…1 bar
(4 bar)
100 kPa
(400 kPa)
15 psig
(60 psig)
3M 0…4 bar
(16 bar)
400 kPa
(1.6 MPa)
60 psig
(240 psig)
3P 0…10 bar
(40 bar)
1 MPa
(4 MPa)
150 psig
(600 psig)
3S 0…40 bar
(160 bar)
4 MPa
(16 MPa)
600 psig
(2400 psig)
3U 0…100 bar
(400 bar)
10 MPa
(40 MPa)
1500 psig (6000 psig)
3Z 0…400 bar
(600 bar)
40 MPa
(60 MPa)
6000 psig (9000 psig)
Absolute pressure
4H 0…1 bar
(4 bar)
100 kPa
(400 kPa)
15 psia
(60 psig)
4M 0…4 bar
(16 bar)
400 kPa
(1.6 MPa)
60 psia
(240 psig)
4P 0…10 bar
(40 bar)
1 MPa
(4 MPa)
150 psia
(600 psig)
4S 0…40 bar
(160 bar)
4 MPa
(16 MPa)
600 psia
(2400 psig)
4U 0…100 bar
(400 bar)
10 MPa
(40 MPa)
1500 psia (6000 psig)
4Z 0…400 bar
(600 bar)
40 MPa
(60 MPa)
6000 psia (9000 psig)
Negative gauge pressure
7H –1…+1 bar
(4 bar)
–100…100 kPa (400 kPa)
–15…15 psig (60 psig)
7M –1…+4 bar
(16 bar)
–100…400 kPa (1.6 MPa)
–15…60 psig (240 psig)
7P –1…+10 bar
(40 bar)
–0.1…1 MPa (4 MPa)
–15…150 psig (600 psig)
Calibration and Technical Units
1 Calibrated nominal value mbar/bar
2 Calibrated nominal value kPa/MPa
3 Calibrated nominal value mm H
2
O/m H
2
O
4 Calibrated nominal value inch H
2
O
5 Calibrated nominal value kgf/cm
2
6 Calibrated nominal value psi
B Calibrated from … to … technical unit … (bar, kPa, psi …)
9 Others
Electronics, Display
A
Analogue 4…20 mA, without display
C Analogue 4…20 mA, with analogue display
H HART 4…20 mA, without display
J
HART 4…20 mA, with digital display
P
PROFIBUS-PA, without display
R
PROFIBUS-PA, with digital display
Accessories/Certification
1
None
2
Bracket for pipe and wall mounting
C
3.1.B Inspection certificate for all parts in contact with the medium made of 1.4435 (AISI 316L)
9
Others
Code for Process Connection see Page 26
Filling Fluid in Diaphragm Seal and Coupling to Cerabar M
Notes on Code H, F, J: Capillary > 1 m with diaphragm seals from DN 50 or 2" only
A Silicone oil, direct coupling
K
1 m (3.28 ft) capillary tubing with high-temperature oil
D Vegetable oil, direct coupling
H
…m capillary tubing with high-temperature oil
E Glycerine, direct coupling
F
…m capillary tubing with vegetable oil
G High-temperature oil, with temperature spacer 100 mm
J
…m capillary tubing with silicone oil
N Fluorolube, grease-free for oxygen service, direct coupling
Material of Diaphragm
Note: Flanges with extension as well as threaded boss in 1.4435 (AISI 316L) available only
1 1.4435 (AISI 316L)
7
PTFE film 0.09 mm on 1.4435 (not for vacuum)
2 Hastelloy C 276
8
PTFE film 0.25 mm on 1.4435 (not for vacuum)
5 Tantalum
9 Others
PMP 48
Product designation
25
Содержание cerabar M PMP 46