Endress+Hauser cerabar M PMP 46 Скачать руководство пользователя страница 12

Design Planning for
the Diaphragm Seal

Diaphragm Seal Filling Fluid

The temperature and pressure of the
process are of critical importance when
selecting the filling fluid for the
diaphragm seal.

,

The suitability of the fluid to meet the
requirements of the medium must also
be considered. For foodstuffs
applications, only physiologically safe
fluids such as vegetable oil or silicone
oil (AK 100) may be used in the
diaphragm seal.

Guidelines for Mounting Capillary
Tubes

The transmitter should generally be
mounted below the tapping point. A
maximum height difference between the
tapping point and transmitter should not
be exceeded. This will otherwise result
in a break in the column of fluid in the
capillary and damage the diaphragm
seal.

Minimum bending radius of capillary
tubing: 100 mm (3.94 in).

In the case of applications in vacuum,
the instrument should be mounted
below the pressure tapping point.

Smallest Recommended Measuring
Span and Diaphragm Diameter

The effects of temperature cause the
diaphragm seal to expand. This in turn
gives rise to an additional temperature
effect on the zero point. When selecting
the diaphragm seal the following points
are to be observed:

The nominal diameter of the
diaphragm seal determines the width
of the diaphragm.

Temperature effects vary inversely
with the width of the diaphragm
diameter.

The largest possible width of diaphragm
should be chosen for small measuring
spans and/or capillaries so that
temperature effects remain within the
nominal range of the application.

Temperature Effects

The temperature coefficients of the
diaphragm seals as stated in the
technical data and dimensions (page 21
onwards) apply to silicone oil
(calibrating tempe20°C/+68°F)
and are determined by the process and
ambient temperatures.
For other fluids, the T

K

value is to be

multiplied by the T

K

correction factor.

For instruments with capillary tubing, the
total temperature coefficient T

K

can be

calculated by adding the T

K

of the

Cerabar M to that of the diaphragm
seal, together with the T

K

of the capillary

tubing.
The T

K

of the capillary tubing is

determined by the ambient temperature:
T

K

per meter for silicone oil:

0.5 mbar/10 K.

Filling fluid
of
diaphragm seal

Code

Permissible
temperature of
medium at
0.05 bar

p

abs

1 bar

(0.73 psi

p

abs

14.5 psi)

Permissible
temperature of
medium at
p

abs

1 bar

(p

abs

14.5 psi)

Maximum
height
difference
at p

abs

1 bar

(p

abs

14.5 psi)

Minimum
permissible
pressure
at +20°C
(+68°F)

Density
[g/cm

3

]

T

K -

Correction
factor

Remarks

Silicone oil
(AK 100)

A, J

–40…+180°C
(–40…+356°F)

–40… +250°C
(–40…+482°F)

max. 7 m
(max. 23 ft)

10 mbar

abs

(0.15 psi)

0,96

1

standard, foodstuffs
applications

High-
temperature oil
(paraffin)

G,
H, K

–10…+200°C
(+14…+392°F)

–10… +350°C
(+14…+662°F)

max. 7 m
(max. 23 ft)

10 mbar

abs

(0.15 psi)

0.81

0.72

Fluorolube FS-5

1)

N

–40…+80°C
(–40…+176°F)

–40…+175°C
(–40…+347°F)

max. 7 m
(max. 23 ft)

10 mbar

abs

(0.15 psi)

1.87

0.91

inert oil e.g. for oxygen
or chlorine

Glycerine

E

+15…+200°C
(+59…+392°F)

max. 4 m
(max. 13.1 ft)

10 mbar

abs

(0.15 psi)

1.26

0.64

foodstuffs applications

Vegetable oil
(Neobee M20)

D, F

–10…+120°C
(+14…+392°F)

–10…+200°C
(+14…+392°F)

max. 7 m
(max. 23 ft)

10 mbar

abs

(0.15 psi)

0.94

1.05

FDA No.:
21CFR172.856

1) Observe operating limits for oxygen service for non-metallic materials.

12

Содержание cerabar M PMP 46

Страница 1: ...measuring ranges up to 400 bar 6000 psi Electronics Analogue cost effective version with short response time especially for fast processes Smart intelligent with versatile operating procedures via HA...

Страница 2: ...d adapter Page 21 DIN 11851 DN 25 DN 40 DN 50 Clamping bracket Page 22 Clamp 1 1 2 Threaded boss Diaphragm seal G Page 27 DIN ISO 228 1 G 1 G 1 G 2 max 400 bar 6000 psi PMP 48 NPT Page 27 ANSI B1 20 1...

Страница 3: ...is available for versions without a local display Electronic Inserts Cerabar M has three electronics versions Analogue 4 20 mA Operation directly at the measuring point with one potentiometer each for...

Страница 4: ...the form of a four digit number The bar graph depicts the current pressure value related to the measuring range Process Connections Process connections are available with all common threads flush mou...

Страница 5: ...auser Supply voltage 11 5 45 V DC or Ex ia 11 5 30 V DC or Cerabar M pressure transmitter with PROFIBUS PA digital communication signal and connection via segment coupler to a PLC or PC using e g the...

Страница 6: ...sure to a resistance bridge The bridge output voltage which is proportional to pressure is then measured and evaluated Advantages For process pressures up to 400 bar 6000 psi Excellent long term stabi...

Страница 7: ...e When calibrating with reference pressure the pressure for lower range value and upper range value must be entered and confirmed by pressing the Zero or Span key twice Press these keys once to displa...

Страница 8: ...ion Using a Handheld Terminal Using the Universal HART Communicator DXR 275 handheld terminal you can set the Cerabar M make checks and use additional functions such as Damping and Calibration without...

Страница 9: ...and actuators including those in explosion hazardous areas to be connected to bus line With PROFIBUS PA two wire looped instruments can be supplied by the sensor with power and digital process inform...

Страница 10: ...lue Diaphragm seals also shift the zero point depending on the orientation of the instrument neutral calibration position max positive zero point shift max negative zero point shift This zero point sh...

Страница 11: ...nd pipe is available for this Material 1 4301 AISI 304 Order No 52001402 It can also be selected as an accessory in the Product Structure Mounting point remote from the measuring point Measuring point...

Страница 12: ...the application Temperature Effects The temperature coefficients of the diaphragm seals as stated in the technical data and dimensions page 21 onwards apply to silicone oil calibrating temperature 20...

Страница 13: ...on the electronic insert without interrupting the process measurement Always connect the screening or ground cable if present to the internal ground terminal of the housing not to terminal 3 1 2 3 Di...

Страница 14: ...ted to the PROFIBUS network using a preterminated cable Versions PM 4 L1 P PM 4 L1 R PM 4 L2 P PM 4 L2 R Connection M12 Plug For hygienic applications or humid operating areas Endress Hauser also offe...

Страница 15: ...instrument Universal HART Communicator DXR 275 PC with operating program e g Commuwin II via Commubox FXA 191 Optional digital display with PROFIBUS PA electronics see page 8 Connection via segment c...

Страница 16: ...ibration rigs for 40 mbar absolute to 100 mbar absolute 0 3 of set span Warm up time Analogue Smart PROFIBUS PA 200 ms 1 s 1 s Rise time T90 time Analogue Smart PROFIBUS PA 60 ms 220 ms 220 ms Settlin...

Страница 17: ...re over pressure limit OPL MWP nameplate x 1 5 The pressure values permitted at higher temperatures can be found in the following standards EN 1092 1 2001 Tab 18 ASME B 16 5a 1998 Tab 2 2 2 F316 ASME...

Страница 18: ...tronics only Overvoltage category II to DIN EN 61010 1 Ripple No effect for 4 20 mA signal up to 5 residual ripple within permissible range With HART Communicator and Commubox max ripple measured on 5...

Страница 19: ...bar 4 bar 100 100 kPa 400 kPa 15 15 psig 60 psig 7M 1 4 bar 16 bar 100 400 kPa 1 6 MPa 15 60 psig 240 psig 7P 1 10 bar 40 bar 0 1 1 MPa 4 MPa 15 150 psig 600 psig Calibration and Technical Units 1 Ca...

Страница 20: ...40 EL SMS thread 2 PN 40 FG RJT thread 1 PN 40 FL RJT thread 2 PN 40 GG ISS thread 1 PN 40 GL ISS thread 2 PN 40 KL DRD flange D 65 mm LL Varivent type N for pipes DN 40 125 Pipe diaphragm seal PB Dai...

Страница 21: ...weight DN PN D f G m dM Ambient Process max A max A mm bar mm mm mm mm mbar 10K mbar 10K bar mbar mm kg mm kg PMP 46 AG 32 40 50 10 Rd 58 x 1 6 21 32 3 4 from 2 0 9 251 5 1 4 256 5 1 7 PMP 46 AH 40 40...

Страница 22: ...mm mbar 10K mbar 10K bar mbar mm kg mm kg PMP 46 DG 25 33 7 38 25 32 40 1 40 50 5 34 3 4 from 0 4 8 234 5 1 0 239 5 1 3 PMP 46 DL 40 51 50 2 40 64 45 1 2 from 0 1 9 242 5 1 2 247 5 1 5 PMP 46 DU 70 76...

Страница 23: ...ent Code Pipe Conical sleeve Groove nut Diaphragm seal Stainless steel housing Aluminium housing Nominal diameter Nominal pressure Diameter Sleeve height Thread Height Diaphragm diameter Temperature c...

Страница 24: ...eal 125 50 66 105 M10 45 3 13 16 15 20 84 Diaphragm seal Varivent Type N and DRD Instru ment Code Neck Flange Bolt holes Diaphragm seal Stainless steel housing Aluminium housing Nominal pressure Exten...

Страница 25: ...r 4 bar 100 100 kPa 400 kPa 15 15 psig 60 psig 7M 1 4 bar 16 bar 100 400 kPa 1 6 MPa 15 60 psig 240 psig 7P 1 10 bar 40 bar 0 1 1 MPa 4 MPa 15 150 psig 600 psig Calibration and Technical Units 1 Calib...

Страница 26: ...mm GK DN 50 PN 10 40 extended diaphragm 100 mm GU DN 80 PN 10 40 extended diaphragm 100 mm JK DN 50 PN 10 40 extended diaphragm 200 mm JU DN 80 PN 10 40 extended diaphragm 200 mm Flanges dimensions to...

Страница 27: ...silicone oil for other oils see page 12 Recommended minimum measuring range Max effects of mounting position see page 10 Height of Cerabar M Total weight Height of Cerabar M Total weight PN d1 d d2 x...

Страница 28: ...ng Aluminium housing Threaded boss Nominal pressure Diameter Height Thread length Height Diameter Across flats Temperature coefficient T K for silicone oil for other oils see page 12 Recommended minim...

Страница 29: ...EB 25 10 40 115 18 68 4 14 85 28 8 8 11 228 2 8 233 3 1 PMP 48 EC 25 64 160 140 24 68 4 18 100 28 8 8 11 234 4 4 239 4 7 PMP 48 ED 25 250 150 28 68 4 22 105 28 8 8 11 238 5 1 243 5 4 PMP 48 EF 25 400...

Страница 30: ...150 2 48 3 6 00 0 75 3 62 4 0 75 4 75 1 77 1 2 15 254 5 4 5 259 5 4 8 PMP 48 MJ 2 150 4 48 3 6 00 0 75 3 62 4 0 75 4 75 1 77 1 2 20 254 5 4 9 259 5 5 2 PMP 48 NJ 2 150 6 48 3 6 00 0 75 3 62 4 0 75 4...

Страница 31: ...31...

Страница 32: ...CV5 05 02 PT1 Endress Hauser GmbH Co KG Instruments International P O Box 2222 D 79574 Weil am Rhein Germany Tel 07621 975 02 Fax 07621 975 345 http www endress com info ii endress com Hauser Endress...

Отзывы: