Proline Prosonic Flow P 500 HART
Safety function
Hauser
15
The installation-specific measured error depends on the installation conditions on site, such
as the nominal diameter, wall thickness, real pipe geometry or medium. The maximum
installation-specific measured error is:
Nominal diameter
Installation-specific error limits
[mm]
[in]
15
½
±16.12 % o.r.
25
1
±10.50 % o.r.
40
1½
±7.17 % o.r.
50
2
±6.22 % o.r.
65
2½
±6.67 % o.r.
80
3
±6.08 % o.r.
100
4
±5.48 % o.r.
150
6
±4.91 % o.r.
200
8
±4.69 % o.r.
300
12
±4.45 % o.r.
400
16
±4.42 % o.r.
500
20
±4.37 % o.r.
600
24
±4.33 % o.r.
1000
40
±4.05 % o.r.
2000
80
±3.92 % o.r.
3000
120
±3.88 % o.r.
4000
160
±3.86 % o.r.
The determined installation-specific measured error covers the following application
scenarios:
Pipe geometry
In accordance with ASME B36.10M-2015
Deviation of the entry for the measuring pipe
outer diameter
Max. ±1 % or min. ±0.5 mm (±0.02 in)
(average tolerance class D2/T3 as per DIN EN ISO 1127)
Deviation of the entry for the measuring pipe
wall thickness
Max. ±10 %
(average tolerance class D2/T3 as per DIN EN ISO 1127)
Deviation of the sensor distance
• DN 15 to 65 (½ to 2½"): max. ±1 mm (±0.04 in)
• DN 80 to 200 (3 to 8"): max. ±2 mm (±0.08 in)
• DN 300 to 4000 (12 to 160"): max. ±1 % of outer diameter
Number of traverses
• DN 25 to 600 (1 to 24"): 2 traverses
• DN 1000 to 4000 (40 to 160"): 1 traverse