Technical Specifi cations (continued)
Connectors and Cables
Return Electrode Receptacle:
A dual pin receptacle on the front of the EndoShield 2. The Return Electrode
Receptacle connects a dual-area patient return electrode to the EndoShield 2.
Return Electrode Plug:
A single pin plug on the rear of the EndoShield 2. The Return Electrode Plug connects
to the electrosurgical generator's patient return electrode receptacle.
AEM Cord Connector:
The AEM Cord Connector connects the EndoShield 2 to an AEM instrument cord.
Foot-Control Plug:
A single active pin that connects the EndoShield 2 to the electrosurgical generator’s
footswitch accessory receptacle.
Maximum Electrosurgical
Generator Voltage
4.1 kV peak.
Electrical Characteristics
Power Source
Lithium Battery, 3V CR2.
Patient Leakage Current
Source or sink leakage current is 10 μA maximum.
Dimensions and Weight
External Dimensions
4” tall x 2” wide x 2.5” deep, excluding integrated cords.
AEM Instrument Connector: 7 inch cord length.
Foot-Control Plug: 6 inch cord length.
Weight
140 g (5 oz).
Environmental Characteristics
Operating Temperature
59° to 104° F (15° to 40° C).
Storage and Transport Temperature
-13° to 140° F (-25° to 60° C).
Operating, Storage and Transport Humidity
5% to 95% relative, non-condensing.
Atmospheric Pressure (Operating)
70 - 110 kPa.