Technical Specifi cations (continued)
Guidance and Manufacturer’s Declaration
–
Immunity
The EndoShield 2 is intended for use in the electromagnetic environment specifi ed below. The
customer or user of the EndoShield 2 should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment
–
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6kV Contact
±8kV Air
±6kV Contact
±8kV Air
Floors should be wood, concrete
or ceramic tile. If fl oors are
synthetic, the r/h should be at
least 30%.
Electrical Fast
Transient (EFT)
IEC 61000-4-4
±2kV Mains
±1kV Input/Output
(I/Os)
N/A
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1kV Diff erential
±2kV Common
N/A
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage Dips/
Dropout
IEC 61000-4-11
>95% Dip for
0.5 Cycle
60% Dip for
5 Cycles
30% Dip for
25 Cycles
>95% Dip for
5 Seconds
N/A
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the EndoShield 2 requires
continued operation during
power mains interruptions,
it is recommended that the
EndoShield 2 be powered from
an uninterruptible power supply
or a battery.
Power
Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic
fi elds should be that of a
typical commercial or hospital
environment.