EM TEST (Switzerland) GmbH
CWS500 N1.3 / N1.4
User manual
V 1.01
6 / 47
2.
Safety
2.1.
Safety Aspects
Tests performed with RF-disturbances, induced by radio-frequency fields, are immunity tests on electronic
equipment. Therefore, it is the responsibility of the user to avoid critical failures and risks to the environment and
the operators.
Long power supply lines to the EUT may radiate energy. It is the responsibility of the user to determine whether
it is allows conducting immunity tests in a given location. The CWS 500N1.x are RF generators and therefore
connected cables can radiate and have interferences into the environment. It is in the responsibility of the user
that sensible areas will not disturbing by the RF tests.
The RF voltages on the center pin of the RF output connector can be hazardous. The RF output connector has
to connect to a load before applying AC power to the CWS 500N1.x. Do not touch the center pin of the RF
output connector or accessories, connected to it.
Switch the equipment to the “Test OFF” position before
disconnecting or connecting the load to the RF output connector.
During this RF test, the interference power can be 80W or more. Therefore, the test should perform in a
shielded room to prevent interference with other equipment in the vicinity. Also, see the information included in
the declaration of conformity for the CE mark.
For the test setup, national and international safety regulations must observed.
Individuals with pacemakers must not allow participating in the testing procedure.
WARNING
Before using the CWS 500 N1.3 / N1.4 read the following manuals carefully :
- Safety requirements manual
- Manual for operation CWS 500N1.x generator