Step 2
Page 3
(Service )
Function check Powermeter
step 2/2
- Connect the MONITOR input to the
RF OUT via a ATT 6 dB attenuator
Continue
break
F1
F2
F3
F4
F5
F6
F7
Fig 6.3 :Setup step 2 with connected 6 dB attenuator
Result after the test
Finally the result of the check appears in the display.
Page 3
(Service)
Function check Powermeter
result
In case of a fail result
Check the T50 terminating resistor and 6dB
attenuator and repeat the test.
contact your local EM TEST representative.
MONITOR input test
Pass
FWD and REV inputs test
Pass