EM TEST (Switzerland) GmbH
CWS500 N1.3 / N1.4
User manual
V 1.01
36 / 47
8.
Setup
8.1.
Test Setup with CDN’s
The following information is important for a correct test setup:
Fig 8.1 :Test setup with CDN
•
Each, the EUT and the AE are
10 cm isolated
above the reference ground surface.
•
All outgoing lines have to be
decoupled via CDN
.
•
The CDNs have to be inserted in a distance of
0.1 m up to 0.3 m
from the EUT.
•
The CDNs have to be connected well
with
the
reference ground plane
in a RF-accordant manner.
•
Only
one
CDN is terminated with
a 50
load resistor.
•
Refer to IEC 61000-4-6 Standard for more information to terminate coupling/decoupling devices
Fig 8.2 :Test Setup using CDN’s