5.5.2. Saturation check
For the saturation check the generator increase to the calibrated signal with +5.1dB for check
Amplifier Saturation Test
V = 10.0 V
CAL = F2
START CHANGE
f = 000.100 MHz
F1
F2
F3
F4
F5
F6
F7
CHANGE
:
F1:
Voltage level of the calibrated signal
F4:
Calibration storage file F1 to F5
START
: Starts the saturation test with the selected level and calibration file.
Display during the saturation test
Amplifier Saturation Test
V = 10.0 V CAL = F2
Amplifier saturation Test in progress
Meas 1 = -28.6 dBm Meas 2 = 23.2 dBm
Diff = +5.4 dB
Break
f = 004.220 MHz
F1
F2
F3
F4
F5
F6
F7
Display at the end of the saturation test
Amplifier Saturation Test
V = 10.0 V
CAL = F2
Amplifier Saturation Test finished
START CHANGE
f = 000.100 MHz
F1
F2
F3
F4
F5
F6
F7