Loop 1
(main menu – settings – more – loop – loop1)
Use this function to adjust the amperage sent to loop1. The second line shows
the current amperage (I=2375). Default setting is 'Standard', which should give
an amperage of approximately 2300 (+/- 100). Choose 'Medium' to get an
amperage of approximately 4700 (+/- 100), and 'High' to get the maximum
amperage, which is approximately 6800 (+/- 100). You may reset the amperage
for loop1 by choosing ‘Reset’.
Loop 2
(main menu – settings – more – loop – loop2)
Use this function to adjust the amperage sent to loop2. The second line shows
the current amperage (I=2375). Default setting is 'Standard', which should give
an amperage of approximately 2300 (+/- 100). Choose 'Medium' to get an
amperage of approximately 4700 (+/- 100), and 'High' to get the maximum
amperage, which is approximately 6800 (+/- 100). You may reset the amperage
for loop1 by choosing ‘Reset’.
Language
(main menu – settings – more – language)
Choose the preferred language for the menus. Default menu choice is English.