INDICATIONS & REMEDIES
PLEASE SEE BELOW FOR EASY INSTRUCTIONS FOR ADDRESSING
ISSUES THAT MAY ARISE WHILE USING THE EPIX360™
TROUBLESHOOTING, SUPPORT, & REPAIR
Issue
Reason
Troubleshooting
No Power to
the System
(Battery)
• Battery Not Installed
Correctly
• Battery Not Charged
• Battery Does Not
Slide Easily into the
slot (Do not try to
slam the battery to
seat)
Verify battery connector
issue by trying the 2nd
battery to determine if it will
not seat correctly. if not,
Battery is mechanically
obstructed
System
operates, but
red indicator
light not
illuminating
• High Voltage power
supply LED is
broken
• High voltage power
supply is not
working
Call Customer Care
No Mist or
Inadequate
Mist
• Tank Empty
• Tank Not Seated
Correctly
• System Lines not
primed
• Battery Not fully
charged
• Nozzle Clogged
• Release check valve
missing from port of
system
• Internal Component
Leak
• Pump Failure
• Fill Tank
• Remove Tank and
Re-apply
• Prime System for 5
to 10 seconds after
replacing tank
• Charge to Full battery
• Remove Nozzle Tip to
check if any residual
or contaminant is
blocking
• Replace check valve if
missing
• Call Customer Care
Nozzle Tip
leaking or
dribbling
• Remove tank when
system not in use.
• Nozzle Extension &
Tip are not lined up
or check valve is not
seated correctly
Verify the nozzle tip is
hand tighened in place
System turns
on and off
automatically
• Trigger button failure
• Main electrical
componets have
been exposed to
liquid/moisture
Call Customer Care