Preparation
Lift the clamp handles to release and open the
toggle clamps.
Remove the cell lid taking care not to damage
the o-ring.
A
Cell lid
B
O-ring
The two cell mirrors can be seen at opposite
ends of the analysis cell as shown.
The cell mirror is recessed back into the ends of
the analysis cell, making it difficult to see when
cleaning.
Note
The cell mirror coating is delicate and highly
reflective. Excessive force
WILL
permanently
damage the mirror coating putting the leak
detector out of operation.
ONLY
trained operators should perform this
task.
Maintaining the leak detector
Reference Manual
July 2022
00809-0100-4245
80
Rosemount CT4215
Содержание Rosemount CT4215