Once the leak detector has been moved into
position the transportation wheels must be
removed.
Note
There is limited clearance between the M8 bolt
head and the floor as shown opposite.
Remove the 8 off M8 bolts and washers securing
the transportation wheels to the leak detector
base.
It will be necessary to support the leak detector
enclosure while the transportation wheels are
slid out from under the unit.
Remove the two press fit covers to access the
bolting position.
Fit the securing bolts loosely to locate the leak
detector in place.
DO NOT TIGHTEN
at this time.
Remove the 4 off press fit covers to access the
adjustable feet.
Using a 14mm AF socket adjust the feet to suit
giving a horizontal stable platform for the leak
detector.
Replace the covers.
Reference Manual
Installation
00809-0100-4245
July 2022
Reference Manual
35
Содержание Rosemount CT4215
Страница 1: ...Reference Manual 00809 0100 4245 Rev AC July 2022 Rosemount CT4215 Leak Detection System ...
Страница 12: ...Contents Reference Manual July 2022 00809 0100 4245 12 Rosemount CT4215 ...
Страница 46: ...Start up procedure Reference Manual July 2022 00809 0100 4245 46 Rosemount CT4215 ...
Страница 76: ...Software interface description Reference Manual July 2022 00809 0100 4245 76 Rosemount CT4215 ...
Страница 86: ...Maintaining the leak detector Reference Manual July 2022 00809 0100 4245 86 Rosemount CT4215 ...
Страница 92: ...Troubleshooting and diagnostics Reference Manual July 2022 00809 0100 4245 92 Rosemount CT4215 ...