A. Pack position when the timer starts.
B. Pack position when the timer stops for the first sample measurement bar.
C. Pack position when the timer stops for the second sample measurement bar.
3. Repeat the measurement a number of times to enure consistent results.
4. Convert the time taken to msecs (milliseconds). One second = 10000 msecs.
5. Enter the time taken in msecs to set the required measurement start time.
6. Repeat this process to measure the time taken for the pack to pass between the
input gate and the second sample measurement bar. Add a further 500 msecs to
the time taken and enter it as the initial measurement end time.
7. A test pack with a known leak situated at the far end of the pack will be sent along
the product line to confirm the setup.
8. Leak detection is confirmed when a spike has been detected on the concentration
screen as shown.
Software interface description
Reference Manual
July 2022
00809-0100-4245
70
Rosemount CT4215
Содержание Rosemount CT4215
Страница 1: ...Reference Manual 00809 0100 4245 Rev AC July 2022 Rosemount CT4215 Leak Detection System ...
Страница 12: ...Contents Reference Manual July 2022 00809 0100 4245 12 Rosemount CT4215 ...
Страница 46: ...Start up procedure Reference Manual July 2022 00809 0100 4245 46 Rosemount CT4215 ...
Страница 76: ...Software interface description Reference Manual July 2022 00809 0100 4245 76 Rosemount CT4215 ...
Страница 86: ...Maintaining the leak detector Reference Manual July 2022 00809 0100 4245 86 Rosemount CT4215 ...
Страница 92: ...Troubleshooting and diagnostics Reference Manual July 2022 00809 0100 4245 92 Rosemount CT4215 ...