These are entered during the creation of the log file.
Date/Time
The time stamp for each unit; this is set by the PC.
Leak (abs)
The maximum concentration value for each unit. If this value is greater
than the upper threshold, this package will be rejected.
Leak (cal)
For certain applications, a calibration factor may be applied in order to
have units for leakage (i.e., bubbles/second or mg/day).
UT
The upper threshold value. If the concentration exceeds this value, the
package is rejected.
LT
The lower threshold: If two consecutive concentration points exceed
this value, the package is rejected.
Noiselevel
The noise value at the time when the package is logged.
Reject
Packages which do not breach thresholds are
reject = 0
, and those
which breach thresholds are
reject =1
.
Note
If you want to view only rejected packages, then filter column
H
to
show only rows with
1
in them.
NOTICE
The
CanInfo
type is logged for all units processed; the
Reject
type only shows the reject
value.
Reference Manual
Software interface description
00809-0100-4245
July 2022
Reference Manual
61
Содержание Rosemount CT4215
Страница 1: ...Reference Manual 00809 0100 4245 Rev AC July 2022 Rosemount CT4215 Leak Detection System ...
Страница 12: ...Contents Reference Manual July 2022 00809 0100 4245 12 Rosemount CT4215 ...
Страница 46: ...Start up procedure Reference Manual July 2022 00809 0100 4245 46 Rosemount CT4215 ...
Страница 76: ...Software interface description Reference Manual July 2022 00809 0100 4245 76 Rosemount CT4215 ...
Страница 86: ...Maintaining the leak detector Reference Manual July 2022 00809 0100 4245 86 Rosemount CT4215 ...
Страница 92: ...Troubleshooting and diagnostics Reference Manual July 2022 00809 0100 4245 92 Rosemount CT4215 ...