Success: Calibration completed successfully.
•
Failed Constant: The calculated calibration constant is outside the range of
±20.00.
•
Failed Slope: The calculated calibration slope is outside the range of 34.5 to 57.5.
•
WarmUp Abort: Attempted to perform a calibration during warmup.
•
Alarm Abort: Another alarm occurred and caused the calibration cycle to abort.
12.
In the event of a calibration failure, do the following:
a. From the RESULT menu, press
HOME
to return to the main menu.
b. From the main menu, select
SERVICE TOOLS
.
c. From the OVERVIEW menu, select
ALERTS
.
d. Frpom the ALERTS menu, select
ACTIVE ALERTS
.
The ACTIVE ALERTS menu should contain an
A: CALIBRATION FAILED
alarm.
e. Press the Left key to return to the ALERTS menu.
f. From the ALERTS menu, select
DEVICE STATUS
.
g. From the DEVICE STATUS menu, select
ACKNOWLEDGE
.
h. From the ACKNOWLEDGE menu, select
ACK CALIBRATION FAILED
.
When the process is complete, the system returns to the ACKNOWLEDGE menu.
3.10.2
Calibration log
The Rosemount 6888A stores the current and 10 previous calibrations. The stored data
includes the slope, constant, cell impendance, and time (in days) since that calibration.
The log can be accessed as follows:
Using the Field Communicator: Configure > Calibration > Calibration Constants >
Calibration Logs.
When first entering the log, the screen displays
Calibration 1
. This is the most recent
previous calibration, not the current calibration.
Calibration 0
displays the current
calibration. Use the keypad or communicator keys to navigate through the logs. The
factory default for log entries is a slope of 50.00 mV/Dec and a constant of 0.00 mV. Any
log entries with these values means there have not been at least 10 calibrations performed
to fill the entire log.
3.10.3
Reset calibration
The Rosemount 6888A can reset the current calibration and replace it with the previous
calibration. You may reset the calibration in cases where the calibration was known to
contain invalid values. Resetting a calibration transfers the slope and constant from
calibration 1 in the log into the calibration 0 position. The cell impedance and days since
calibration remain unchanged.
Configuration, startup, and operation
Reference Manual
83
Содержание Rosemount 6888A
Страница 1: ...Reference Manual 00809 0100 4890 Rev AA July 2018 Rosemount 6888A O2 Combustion Flue Gas Transmitter ...
Страница 4: ......
Страница 8: ...Contents iv Rosemount 6888A ...
Страница 61: ...Calibration Defaults Figure 3 1 Configuration startup and operation Reference Manual 53 ...
Страница 62: ...HART Menu Trees Rosemount 6888Xi Figure 3 2 Configuration startup and operation 54 Rosemount 6888A ...
Страница 63: ...Configuration startup and operation Reference Manual 55 ...
Страница 64: ...Configuration startup and operation 56 Rosemount 6888A ...
Страница 65: ...Configuration startup and operation Reference Manual 57 ...
Страница 66: ...Configuration startup and operation 58 Rosemount 6888A ...
Страница 67: ...Configuration startup and operation Reference Manual 59 ...
Страница 68: ...Configuration startup and operation 60 Rosemount 6888A ...
Страница 69: ...HART Menu Trees 375 475 Field Communicator Figure 3 3 Configuration startup and operation Reference Manual 61 ...
Страница 70: ...Configuration startup and operation 62 Rosemount 6888A ...
Страница 71: ...Configuration startup and operation Reference Manual 63 ...
Страница 72: ...Configuration startup and operation 64 Rosemount 6888A ...
Страница 74: ...Configuration startup and operation 66 Rosemount 6888A ...
Страница 75: ...Configuration startup and operation Reference Manual 67 ...
Страница 76: ...Configuration startup and operation 68 Rosemount 6888A ...
Страница 77: ...Configuration startup and operation Reference Manual 69 ...
Страница 79: ...Configuration startup and operation Reference Manual 71 ...
Страница 80: ...Configuration startup and operation 72 Rosemount 6888A ...
Страница 81: ...Configuration startup and operation Reference Manual 73 ...
Страница 82: ...Configuration startup and operation 74 Rosemount 6888A ...
Страница 93: ...D A Trim Method Simplified Figure 3 7 Configuration startup and operation Reference Manual 85 ...
Страница 94: ...Configuration startup and operation 86 Rosemount 6888A ...
Страница 104: ...Troubleshooting 96 Rosemount 6888A ...
Страница 120: ...Maintenance and service 112 Rosemount 6888A ...
Страница 128: ...Optional accessories 120 Rosemount 6888A ...
Страница 132: ...Rosemount 6888 product certifications 124 Rosemount 6888A ...
Страница 135: ...Safety data Reference Manual 127 ...