3
Configuration, startup, and operation
WARNING!
ELECTRIC SHOCK
Install all protective equipment covers and safety ground leads before equipment startup.
Failure to install covers and ground leads could result in serious injury or death.
CAUTION!
EQUIPMENT DAMAGE
If external loop power is used, the power supply must be a safety extra low voltage (SELV)
type.
3.1
Power up Rosemount 6888 Transmitter
without Rosemount 6888Xi
Complete the following steps to apply power to the Rosemount 6888 Transmitter without
connecting it to the Rosemount 6888Xi.
1.
Apply AC line power to the transmitter.
2.
Apply 24 Vdc loop power to the transmitter.
3.
Using either the DCS control or a Field Communicator, verify communications to the
transmitter.
The transmitter probe takes approximately 45 minutes to warm up to the 736 °C
(1357 °F) heater setpoint. The 4-20 mA signal remains at a default value of 3.5 mA,
and the O
2
reading remains at 0% through this warm-up period. After warm-up, the
probe begins reading oxygen, and the 4-20 mA output is based on the default range
of 0-10% O
2
.
If there is an error condition at startup, an alarm message is displayed.
Configuration, startup, and operation
Reference Manual
47
Содержание Rosemount 6888A
Страница 1: ...Reference Manual 00809 0100 4890 Rev AA July 2018 Rosemount 6888A O2 Combustion Flue Gas Transmitter ...
Страница 4: ......
Страница 8: ...Contents iv Rosemount 6888A ...
Страница 61: ...Calibration Defaults Figure 3 1 Configuration startup and operation Reference Manual 53 ...
Страница 62: ...HART Menu Trees Rosemount 6888Xi Figure 3 2 Configuration startup and operation 54 Rosemount 6888A ...
Страница 63: ...Configuration startup and operation Reference Manual 55 ...
Страница 64: ...Configuration startup and operation 56 Rosemount 6888A ...
Страница 65: ...Configuration startup and operation Reference Manual 57 ...
Страница 66: ...Configuration startup and operation 58 Rosemount 6888A ...
Страница 67: ...Configuration startup and operation Reference Manual 59 ...
Страница 68: ...Configuration startup and operation 60 Rosemount 6888A ...
Страница 69: ...HART Menu Trees 375 475 Field Communicator Figure 3 3 Configuration startup and operation Reference Manual 61 ...
Страница 70: ...Configuration startup and operation 62 Rosemount 6888A ...
Страница 71: ...Configuration startup and operation Reference Manual 63 ...
Страница 72: ...Configuration startup and operation 64 Rosemount 6888A ...
Страница 74: ...Configuration startup and operation 66 Rosemount 6888A ...
Страница 75: ...Configuration startup and operation Reference Manual 67 ...
Страница 76: ...Configuration startup and operation 68 Rosemount 6888A ...
Страница 77: ...Configuration startup and operation Reference Manual 69 ...
Страница 79: ...Configuration startup and operation Reference Manual 71 ...
Страница 80: ...Configuration startup and operation 72 Rosemount 6888A ...
Страница 81: ...Configuration startup and operation Reference Manual 73 ...
Страница 82: ...Configuration startup and operation 74 Rosemount 6888A ...
Страница 93: ...D A Trim Method Simplified Figure 3 7 Configuration startup and operation Reference Manual 85 ...
Страница 94: ...Configuration startup and operation 86 Rosemount 6888A ...
Страница 104: ...Troubleshooting 96 Rosemount 6888A ...
Страница 120: ...Maintenance and service 112 Rosemount 6888A ...
Страница 128: ...Optional accessories 120 Rosemount 6888A ...
Страница 132: ...Rosemount 6888 product certifications 124 Rosemount 6888A ...
Страница 135: ...Safety data Reference Manual 127 ...