System Parameters
(continued)
Table 3-1:
Parameter name
Unit
Description
Cal Result
--
Calibration result:
0
= None
1
= Success
2
= Failed constant
3
= Failed slope
4
= Failed temperature
5
= Gas 1 tolerance error
6
= Gas 2 tolerance error
7
-
10
(future)
11
= AutoCal no resp
12
= AutoCal out of sync
13
= AutoCal abort
14
= No solenoid
15
= Warmup abort
16
= Alarm abort
3.9
Parameter setup
3.9.1
Test gas values
Use a Field Communicator to set test gas values for calibration. A Rosemount 6888A
shipped from the factory has test gas values for low and high set to 0.4% and 8.0%
respectively. Perform this same process any time a transmitter board is replaced.
1.
Use the 375/475 Field Communicator to access the main HART menu.
2.
From the main menu, select
CONFIGURE
.
3.
From the CONFIGURE menu, select
MANUAL SETUP
.
4.
From the MANUAL SETUP menu, select
CALIBRATION
.
5.
From the CALIBRATION menu, select
GAS 1
.
6.
Press the Right key to edit the gas value.
7.
Use the stylus to enter the new gas value. Then press ENTER when finished.
8.
Use the stylus to select SEND to update the Rosemount 6888A.
9.
Repeat steps 5 through 8 for Gas 2.
3.9.2
Set test gas times
Use a Field Communicator to set test gas flow time and purge time for calibration.
Configuration, startup, and operation
Reference Manual
79
Содержание Rosemount 6888A
Страница 1: ...Reference Manual 00809 0100 4890 Rev AA July 2018 Rosemount 6888A O2 Combustion Flue Gas Transmitter ...
Страница 4: ......
Страница 8: ...Contents iv Rosemount 6888A ...
Страница 61: ...Calibration Defaults Figure 3 1 Configuration startup and operation Reference Manual 53 ...
Страница 62: ...HART Menu Trees Rosemount 6888Xi Figure 3 2 Configuration startup and operation 54 Rosemount 6888A ...
Страница 63: ...Configuration startup and operation Reference Manual 55 ...
Страница 64: ...Configuration startup and operation 56 Rosemount 6888A ...
Страница 65: ...Configuration startup and operation Reference Manual 57 ...
Страница 66: ...Configuration startup and operation 58 Rosemount 6888A ...
Страница 67: ...Configuration startup and operation Reference Manual 59 ...
Страница 68: ...Configuration startup and operation 60 Rosemount 6888A ...
Страница 69: ...HART Menu Trees 375 475 Field Communicator Figure 3 3 Configuration startup and operation Reference Manual 61 ...
Страница 70: ...Configuration startup and operation 62 Rosemount 6888A ...
Страница 71: ...Configuration startup and operation Reference Manual 63 ...
Страница 72: ...Configuration startup and operation 64 Rosemount 6888A ...
Страница 74: ...Configuration startup and operation 66 Rosemount 6888A ...
Страница 75: ...Configuration startup and operation Reference Manual 67 ...
Страница 76: ...Configuration startup and operation 68 Rosemount 6888A ...
Страница 77: ...Configuration startup and operation Reference Manual 69 ...
Страница 79: ...Configuration startup and operation Reference Manual 71 ...
Страница 80: ...Configuration startup and operation 72 Rosemount 6888A ...
Страница 81: ...Configuration startup and operation Reference Manual 73 ...
Страница 82: ...Configuration startup and operation 74 Rosemount 6888A ...
Страница 93: ...D A Trim Method Simplified Figure 3 7 Configuration startup and operation Reference Manual 85 ...
Страница 94: ...Configuration startup and operation 86 Rosemount 6888A ...
Страница 104: ...Troubleshooting 96 Rosemount 6888A ...
Страница 120: ...Maintenance and service 112 Rosemount 6888A ...
Страница 128: ...Optional accessories 120 Rosemount 6888A ...
Страница 132: ...Rosemount 6888 product certifications 124 Rosemount 6888A ...
Страница 135: ...Safety data Reference Manual 127 ...